RECORD AN ECG
20
For accurate V-lead placement and monitoring, it is important to locate the 4
th
intercostal space. The 4
th
intercostal
space is determined by first locating the 1
st
intercostal space. Because patients vary with respect to body shape, it is
difficult to palpate the 1
st
intercostal space with accuracy. Thus, locate the 2
nd
intercostal space by first palpating the
little bony prominence called the
Angle of Louis
, where the body of the sternum joins the manubrium. This rise
in the sternum identifies where the second rib is attached, and the space just below it is the 2
nd
intercostal space.
Palpate and count down the chest until you locate the 4
th
intercostal space.
Patient Hookup Summary Table
AAMI Lead
IEC
Lead
Electrode
Position
Red
Red
On the 4
th
intercostal space
at the right sternal
border.
Yellow
Yellow
On the 4
th
intercostal space
at the left sternal
border.
Green
Green
Midway between
V2/C2 and V4/C4
electrodes.
Blue
Brown
On the 5
th
intercostal space
at the left
midclavicular line.
Orange
Black
Midway between
V4/C4 and V6/C6
electrodes.
Violet
Violet
On the left
midaxillary line,
horizontal with
V4/C4 electrode.
Black
White
Yellow
Red
On the deltoid,
forearm, or wrist.
Red
Green
Green
Black
On the thigh or
ankle.