13
GENERAL CARE
Precautions
Turn off the device before inspecting or cleaning.
Do not immerse the device in water.
Do not use organic solvents, ammonia-based solutions, or abrasive cleaning agents which may damage
equipment surfaces.
Inspection
Inspect your equipment daily prior to operation. If you notice anything that requires repair, contact an authorized
service person to make the repairs.
Verify that all cords and connectors are securely seated.
Check the case and chassis for any visible damage.
Inspect cords and connectors for any visible damage.
Inspect keys and controls for proper function and appearance.
Cleaning and Disinfecting
1.
Disconnect the power source. Remove cables and lead wires from device before cleaning.
2.
For general cleaning of cables and lead wires, use a soft, lint-free cloth lightly moistened with a mild soap
and water solution. Wipe and air dry.
3.
For disinfecting the exterior surfaces of the device, patient acquisition module, cables and lead wires, wipe
exterior using:
Clorox Healthcare
®
Bleach Germicidal Wipes (use according to instructions on product label),
or
a soft, lint-free cloth with a solution of Sodium Hypochlorite (10% household bleach and water solution)
minimum 1:500 dilution (minimum 100 ppm free chlorine) and maximum 1:10 dilution as recommended
by the APIC Guidelines for Selection and Use of Disinfectants.
WARNING
:
Prevent liquid from penetrating the device and do not attempt to clean/disinfect the device or patient
cables by submerging into a liquid, autoclaving, or steam cleaning.
Do not expose cables to strong ultra-violet radiation.
Do not sterilize the device or lead wires with Ethylene Oxide (EtO) gas.
Do not immerse cable ends or lead wires; immersion can cause metal corrosion. Use caution with excess
liquid as contact with metal parts may cause corrosion.
Do not use excessive drying techniques such as forced heat.
Improper cleaning products and processes can damage the device, produce brittle lead wires and cables,
corrode the metal, and void the warranty. Use care and proper procedure whenever cleaning or
maintaining the device.