SYSTEM SETTINGS
NOTE
: The plot-frequency filter does not filter the digitized signal acquired for interpretation of the ECG.
NOTE
: The Filter setting can also be changed for a single ECG by touching anywhere on the ECG
waveform in the real-time or acquired view.
Interp
This selection allows the clinician to turn the ECG interpretation statements On/Off for presentation on the display
and/or plotted report.
NOTE:
The ECG interpretations offered by the device are intended to be most relevant when used in
conjunction with a physician over-read and with consideration of all other relevant patient data.
Reasons
This selection allows the clinician to include additional information about the ECG interpretation on the printout.
Reasons statements are specific details that indicate why a particular interpretive statement was chosen. Reasons
statements print in brackets [ ] within the interpretive text. Reasons are only available if the interpretation statements
option is enabled. Turning the reasons statement function ON or OFF does not affect the measurements criteria or
the interpretive statement selected by the analysis program.
For Example:
Anteroseptal Infarct [40+ ms Q WAVE IN V1-V4]
“Anteroseptal Infarct” is the interpretive statement.
“40+ ms Q WAVE IN V1-V4” is the reason statement explaining the interpretive statement.
Append
This selection allows the user to add a status message or statement phrase to the ECG to be printed under the
interpretive text printout. The choices are “UNCONFIRMED REPORT” or “Reviewed by.”
Delete Rule
The Delete Rule allows the user to define how and when ECGs are automatically marked for deletion in the ECG
directory. ECGs can be marked manually or marked for deletion per the Delete Rule. ECG records are only fully
deleted from the directory when the directory becomes full. When the memory is full, ECGs that are marked for
deletion will be automatically deleted based on their acquisition date and time (first-in/first-out) to make room for
new ECG records. More than one ECG record may be removed from the directory to make room for the new
incoming record.
The Delete Rule selections are:
1.
Post Plot = ECG is automatically marked for deletion after printing
2.
Post Transmit = ECG is automatically marked for deletion after transmission
3.
Post Plot/Transmit = ECG is automatically marked for deletion after transmission and printing
NOTE:
A record may be permanently deleted by selecting the record, followed by selecting Erase from the
left side of the display. A window will be presented as “Erase ECG?” If you select Yes the record will be
permanently erased. If you select No the record will be maintained in the file.