Operation  2018-10-21

 

Page2

 

 

 

This equipmentt must not be connected to or used in combination with any other apparatus or system.   
It must not be used as an integral component of any other apparatus or system. 
J. MORITA MFG. CORP. will not be responsible for accidents, equipment damage, bodily injury or 
any other trouble which results from ignoring this prohibition.   

 

Accurate canal measurement is not always possible depending on the shape and condition of the 
tooth as well as a decline in the equipment’s performance. 

 

Do not use damaged file holders; an accurate measurement can not be made with a damaged file 
holder. 

 

When continuous tone is heard while the main power switch is on and without any operation, some 
electrical part may be malfunction. Do not use the equipment and send it to J. MORITA OFFICE for 
repairing. 

 

A rubber dam should be used when performing endodontic treatment. 

 

Caution: US Federal law restricts this equipment to sale by or on the order of a dentist in U.S.A. 

 

The DENTAPORT ZX needs special precaution regarding EMC and needs to be installed and put 
into service according to the EMC information provided in the Accompanying Documents. 

 

Portable and mobile RF communications equipment can affect the DENTAPORT ZX. 

 

Use of the parts other than those accompanied or specified by J. MORITA MFG. CORP. may result 
in increased EMC emissions or decreased EMC immunity of the DENTAPORT ZX. 

 

The DENTAPORT ZX should not be used adjacent to or stacked with other equipment and that if 
adjacent or stacked use is necessary, the DENTAPORT ZX should be observed to verify normal 
operation in the configuration in which it will be used. 

 

No modification of this equipment is allowed. 

 

 
 

 

 

Do not use this equipment in conjunction with an electric scalpel or on patients who have a 
pacemaker. 

 

Do not use this equipment in the medical operation room. 

 

Blocked canals cannot be accurately measured. 

 

This equipment must not be connected to or used in combination with any other apparatus or system.   
It must not be used as an integral component of any other apparatus or system. 
J. MORITA MFG. CORP. will not be responsible for accidents, equipment damage, bodily injury or 
any other trouble which results from ignoring this prohibition. 

 

Illumination devices such as fluorescent lights and the Film viewer which use an inverter can cause 
the DENTAPORT ZX to operate erratically.    Do not use the DENTAPORT ZX near devices such as 
these. 

 

Electromagnetic wave interference could cause this device to operate in an abnormal, random and 
possibly dangerous manner.    Cellular phone, transceivers, remote controls and all other devices 
which transmit electromagnetic waves located inside the building should be turned off. 

 

Do not perform maintenance while using the instrument for treatment. 

 

 

 

 

 

 

 

Содержание DENTA PORT ZX

Страница 1: ...Instructions This is the Canal Measurement Module The Canal Preparation and Light Cure Module sold separately can be easily connected to this module so that preparation can be performed while measuri...

Страница 2: ......

Страница 3: ...ents other than those specified by J MORITA MFG CORP and other than in their original condition 5 Operating the equipment in ways other than the operating procedures described in this manual or result...

Страница 4: ...y 9 Meter Display 10 Operating the Unit 11 Root Canals not suitable for Electronic Measurement 13 EMR and Radiography 15 6 After Using the Unit 16 7 Maintenance 18 Autoclavable Components 18 Non Autoc...

Страница 5: ...equipment itself or causing bodily injury Note the meaning of the following symbols and expressions This warns that it may result serious injury of the patient or operator if the instructions are not...

Страница 6: ...r specified by J MORITA MFG CORP may result in increased EMC emissions or decreased EMC immunity of the DENTAPORT ZX The DENTAPORT ZX should not be used adjacent to or stacked with other equipment and...

Страница 7: ...Operator Profile This equipment must only be used by dentists and other legally licensed professionals Patient Population Age Child to Elderly Weight N A Nationality N A Sex N A Health It is not inten...

Страница 8: ...File Holder Contrary Electrode Lip Clip Probe Cord Canal Measurement Module Power Switch 2 Parts Identification Accessories Probe Cord File Holder Contrary Electrode Code No 7503661 Code No 7503670 Co...

Страница 9: ...e plus end down into place 2 Make sure the contacts are not bent or damaged Do not reverse the plus and minus poles Never allow the spring contact to push against the edge of the battery This could da...

Страница 10: ...the white female connector on the probe cord Make sure to match colors of the file holder and contrary electrode to the probe cord Measurements cannot be made if these connections are reversed Checki...

Страница 11: ...t the tester into the probe cord jack Check that the meter indicates within 3 bars away from above or below 1 The meter may jump when the tester is inserted If it does wait for about one second until...

Страница 12: ...hing Never use the unit if the battery power indicator is flashing on and off The unit will not function properly if the battery power is low Battery power indicator also appears when the Canal Prepar...

Страница 13: ...lected The apical line can be set anywhere between 2 and Apex All memory settings will be retained even after the unit is turned off Simply select M1 M2 or M3 to use those memory settings Check the se...

Страница 14: ...is advanced down towards the apex In some cases such as a blocked canal a measurement cannot be made For details see Root Canals not suitable for Electronic Measurement on page 13 Always check the me...

Страница 15: ...patient about this before using the contrary electrode Take care that medicinal solutions such as formalin cresol FC or sodium hypochlorite do not get on the contrary electrode or the file holder The...

Страница 16: ...es except the power switch will work 5 Insert the file in most case size 10 until the meter reads 0 5 this point can be recognized by the change in the alarm sound as well Then advance the file with s...

Страница 17: ...and opening of the canal thoroughly to get rid of all blood saliva and chemical solutions and then make a measurement Broken crown If the crown is broken and a section of the gingival tissue intrudes...

Страница 18: ...tooth Thoroughly remove all the pulp inside the canal otherwise an accurate measurement cannot be made Caries touching the gums In this case electrical leakage through the caries infected area to the...

Страница 19: ...ramen and anatomical apex do not correspond exactly The actual apical foramen may be located up towards the crown In these cases the X ray image will seem to indicate that the file has not reached the...

Страница 20: ...p the probe cord around the body of the unit Replacing Batteries Replace the batteries as soon as the battery power indicator starts flashing To be on the safe side replace the batteries when the batt...

Страница 21: ...k in its original place after attachment push it in the direction shown by the arrow in the illustration After installation give the cover a light tug to make sure it is securely attached Always use a...

Страница 22: ...contrary electrode from the probe cord 2 Clean them off in running water with a soft brush and then wipe off the water If a medical agent being used for the treatment has adhered to the components was...

Страница 23: ...dicinal agents glutaral etc or any other special types of water or commercial cleaning liquids Such liquids may result in plastic degradation metal corrosion and adhesion of the residual medical agent...

Страница 24: ...d after each patient s treatment has been completed The file holder long file holder and contrary electrode are extremely hot after autoclaving do not touch until they cool off Do not sterilize the co...

Страница 25: ...omponents in or wipe it with any of the following functional water acidic electrolyzed water strong alkaline solution and ozone water medical agents glutaral etc medicinal agents glutaral etc or any o...

Страница 26: ...with water Do not use excessive amounts of detergent or water and do not soak the components Do not use paint thinner benzine or similar solutions to clean the components Avoid spilling chemical solut...

Страница 27: ...Inspection This equipment should be inspected every 6 months in accordance with the following maintenance and inspection items Maintenance and Inspection Items 1 Check that the Power switch turns the...

Страница 28: ...ode makes good contact with the oral mucosa Clean the file holder with Ethanol for Disinfection Ethanol 70 vol to 80 vol Canal Length Indicator overreacts or is too sensitive Measurements are too shor...

Страница 29: ...at all or only when the file tip is close to the apical foramen Is the canal blocked Is the apical foramen very large and open Is the canal extremely dry Open the passage all the way through the apica...

Страница 30: ...by measuring at two frequencies and the position of the treatment instrument in the root canal is detected Degree of Protection IEC 60529 IPX0 Protection against Electric Shock Internal powered ME eq...

Страница 31: ...ragile Temperature limitation Atmospheric pressure limitation Humidity limitation Disposal The battery should be recycled Metal parts of the equipment are disposed as scrap metal Synthetic materials e...

Страница 32: ...its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Radiated disturbance CISPR 11 Group 1 Class B This device is suitable for use in all establis...

Страница 33: ...t 2 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply lines IEC 61000 4 11...

Страница 34: ...rked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorp...

Страница 35: ......