Page23
Operation 2018-10-21
8.
Replacement Parts, Transport and Storage Environments
Replacement Parts
* Replace parts as necessary depending on degree of wear and length of use.
* Order replacement parts from your local dealer or J. MORITA OFFICE.
Transport and Storage Environments
Temperature: -10°C to +45°C (+14°F to +113°F)
Humidity: 10% to 85% (without condensation)
Atmospheric Pressure: 70 kPa to 106 kPa
* Store the unit where it will not be exposed to X-rays or direct sunlight.
* If the unit has not been used for a long time, make sure it works properly before use.
* Always remove the batteries prior to storing or shipping the unit.
9.
Inspection
Regular Inspection
* This equipment should be inspected every 6 months in accordance with the following maintenance and
inspection items.
Maintenance and Inspection Items
1.
Check that the Power switch turns the equipment on and off properly.
2.
Insert the Tester and check that the indicator is within ±3 lines of 1 on the meter.
3.
Check that the Mode switch changes the memory from M1 to M2 to M3 etc.
4.
Check that the Select and Set switches work properly.
5.
Check that the probe cord can be properly plugged into its jack.
6.
Check that the file holder's plug can be connected properly to the probe cord and that the file holder can
be clipped onto a file. Check the contrary electrode can be plugged into its probe cord connector.
Parts Lists
Component Description
When
Probe Cord
Probe Cord Assembly
Defective conductivity
File Holder
Contrary Electrode