Troubleshooting
44
SkanWasher 300 User Guide
Application Tips: SkanWasher 300/ SkanStacker
Instrument Functions
Air Pressure
Internal pressure pump creates air pressure within the liquid reservoir.
Air pressure increases the flow rate and has no effect on volume.
Increase the air pressure for higher pressure, decrease for lower
pressure.
Internal pressure pump adjusted through
Edit Program
through the
Setup
Menu
.
Vacuum pressure
Vacuum pressure is regulated from pump gauge.
Vacuum pressure determines the rate of aspiration.
Increasing the vacuum pressure value decreases the vacuum.
To convert mm Hg to mbar pressure consult conversion table at web site:
http://[email protected].
Suggested setting
These values are suggested as a starting point for your assay.
Test between
adjustments to optimize air and vacuum pressure for each specific assay.
External Software Setting
Applications
Vacuum
pressure
setting at
pump
Air Pressure
Range 0.0-0.6
bar
(
Edit Program)
Set Aspirate
Position
Range 0-6 mm
(
Edit Program)
ELISA and
adherent cells
250 mbar
0.25 bar
0-2 mm
(0=bottom of
plate)
Cells
350 mbar
0.15-0.20 bar
2-4 mm
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......