Maintenance
SkanWasher 300 User Guide
31
Rinse Outlet Channels
If it is determined that the aspirate probes (longer probes) are clogged, follow
this procedure after performing the
Clogged Probes
procedure to dislodge
particles.
Step 1
Open the outlet channel seal block to get access to the wash head
outlet channel (see Figure 12).
Remove two hex wrench screws using the hex wrenches available
in the tool kit.
Pull out the wash head block to expose the outlet channels.
Step 2
Observe for particles in outlet channel.
Step 3
If particles not removed, and one probe can be identified to be
clogged, replace probe.
Or
If particles not removed, please consult Technical Support to have
outlet channel replaced.
Figure 12: Wash Head Outlet Channel Seal Block
Removal of wash head
Step 1
Perform
Empty Wash Head Procedure
to drain wash head.
Step 2
Turn instrument off.
Step 3
Unscrew the wash head locking screws (2).
Step 4
Lift the wash head up slowly. Disconnect the air pressure supply
tube and the ribbon cable from the wash head.
Outlet/Aspirate channels
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......