Maintenance
SkanWasher 300 User Guide
37
Aspiration Probe horizontal adjustment.
To be performed when the aspiration probes do not enter the wells or are posi-
tioned incorrectly at the bottom of the well. Verify correct plate orientation on
the microplate platform is used during washing.
Step 1
Remove washhead.
Perform
Empty Wash Head Procedure
to drain wash head.
Turn instrument off.
Unscrew the wash head locking screws (2).
Lift the wash head up slowly. Disconnect the air pressure supply
tube and the ribbon cable from the wash head.
Step 2
Locate the two rods that hold the microplate platform.
Step 3
Loosen the two set-screws that secure the two rods to the plate lift
mechanism.
Step 4
Adjust rods in or out as necessary.
Step 5
Fasten both screws.
Step 6
Remount wash head and verify positioning of probes.
Step 7
If necessary, re-adjust.
Figure 15: Version B Microplate Lift Adjustment