Maintenance
26
SkanWasher 300 User Guide
Weekly Maintenance
Perform maintenance once a week or after use of a contaminated solution in
instrument to decontaminate system as well as clean probes.
Step 1
Perform Daily Maintenance to rinse system.
Step 2
Fill a reservoir bottle with 2% bleach solution or other decontami-
nant solution (20% ETOH or an alkali solution of 5-10% NaOH are
acceptable).
Step 3
Connect reservoir bottle to instrument.
Replace Rinse reservoir bottle with bottle filled with decontami-
nate solution.
Connect tubing from reservoir bottle to
Rinse
port and pressure
tubing to pressure port.
Verify waste tubing is connected to vacuum port and vacuum
pump.
Step 4
Insert a clean microplate into the microplate holder.
S
CAUTION:
Consult chemical
resistance chart to verify that
decontaminant solution will not
harm materials used in the
design of the washer. Solution
that should be avoided: most
acids, acetone, anilin, benzene,
phosphoric acid, sulfuric acid,
xylene, toulene, styrene, and
pyridine.
!! IMPORTANT:
Perform
Daily
Maintenance
before running a
decontaminant solution through
instrument to rinse out all buffer
products.
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......