Dansensor
®
Lippke
VC1400
EN
User Guide
79
6. The
Vacuum Test
screen shows the currently set test vacuum.
To set another test pressure, press
Test vacuum
...
...then enter the required test vacuum (ex. 600 mbar) and press
.
7. Press the
button to start the test.
8. The test starts by evacuating the system until the set vacuum is reached.
Then the valves are closed and in the following period the pressure should be observed
until it can be decided whether the system is tight or not.
The test can then be stopped by pressing the
button.
9. When a test fails, it will be necessary to investigate where there are leaks in the system and
have these repaired.
NOTE! We recommend to perform additional tests afterwards.
Содержание AMETEK Dansensor Lippke VC1400
Страница 1: ...P N 390262 B 02 2022 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 8: ...6 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 20: ...18 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 26: ...24 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 52: ...50 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 56: ...54 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 92: ...90 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 96: ...94 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 98: ...96 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 103: ...This blank page has been inserted to enable double sided printing of the document...
Страница 104: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...