Dansensor
®
Lippke
VC1400
EN
User Guide
23
Bubble test settings
A
Bubble
test is defined using the following parameters:
Parameters marked with an asterisk (*) are the only parameters that can be changed, when test
definition has already been used for testing and if the parameter
Settings ->
Device
->
Lock
Test
has been enabled.
Name
A unique test definition name (up to 30 characters).
Product ID
Unique test definition identification code (up to 20
characters).
Standard Test
*
If set to
No
, the test definition is not available for the
Standard Test Operator
.
Test definition will still be available for an
Operator
and
Supervisor
.
State
Only used if
Use State Mark
is enabled in
Settings ->
Device.
Can be used by a
Supervisor
to categorize the various
test definitions in the test definition list:
New
Automatically assigned to new test
definitions.
Can for example be used for test definitions
where settings are still being considered/
tested.
Current
Typically used for test definitions that have
been approved for use.
This category of test definitions are the only
test definitions available to the
Standard
Test Operator
(provided that the
Standard
Test
option is set to
Yes
).
Old
Can for example be used for discontinued
test definitions or test definitions that
should not be used for a period.
Use ASTM increase vacuum rate
limit
If set to
Yes
the build-up of vacuum in the desiccator is
limited to 30 mbar/sec.
If set to
No
the build-up of vacuum may happen faster,
but it will depend on the capacity of e.g. the pump or
the external vacuum system and how large a volume is
to be evacuated.
Test Vacuum (delta) (V)
The vacuum level to which the desiccator is evacuated.
Test Time (T)
The period of time for which the test vacuum is
maintained in the desiccator.
Custom Fields
*
Select if one or more of the 4 custom fields are required
for the current workflow - see
for details.
Содержание AMETEK Dansensor Lippke VC1400
Страница 1: ...P N 390262 B 02 2022 UserGuide...
Страница 2: ...This blank page has been inserted to enable double sided printing of the document...
Страница 4: ...2 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 8: ...6 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 20: ...18 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 26: ...24 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 52: ...50 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 56: ...54 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 92: ...90 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 96: ...94 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 98: ...96 User Guide EN Dansensor Lippke VC1400 P N 390262 B 02 2022...
Страница 103: ...This blank page has been inserted to enable double sided printing of the document...
Страница 104: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...