EN
OPERATOR'S MANUAL
83
ERRORS (CATEGORY A)
CODE
POSSIBLE CAUSE
SOLUTION
A958
Power supply voltage:
<100Vac;
ranging between 150Vac and 180Vac;
> 270Vac.
Contact your trustworthy plumber to
check the system.
If the problem persists, contact Technical
Service Department.
ERRORS (CATEGORY E)
CODE
POSSIBLE CAUSE
SOLUTION
E000
Interruption of power supply during the
cycle.
Check the continuous presence of power
supply.
If the blackout continues and the
instruments need to be removed, it is
possible to manually open the door as
shown in the figure:
Power plug disconnected during the
cycle.
Check that the plug is correctly plugged
in.
Main switch turned to OFF during the
cycle.
Do not turn off the machine during a
cycle.
E001
Voltage too high in power supply system.
Contact your trustworthy plumber to
check the system.
E004
No network voltage frequency detection.
Contact Technical Service Department.
E011
Attempt to start a cycle with lid open.
Close the lid.
E552
Attempt to start a cycle with chamber
drainage in progress after full chamber
alarm.
Wait for the end of chamber drainage
phase after full chamber alarm.
E070
Lack of salt in the suitable container.
Failure to salt level sensor.
Add salt (see 'Adding salt' section). If the
problem persists, contact Technical
Service Department.
E071
Signal to start a cycle with resin
regeneration as the number of performed
cycles without regeneration has been
reached
Start a cycle with resin regeneration
(Resin regeneration, Disinf. 90°C,
Washing, D1 Custom, D2 Custom and
W1 Custom)
E800
Lack of sufficient amount of detergent to
perform a cycle.
Failure to detergent level sensor.
Add detergent (see 'Detergent reservoir
filling' section).
If the problem persists, contact Technical
Service Department.
E900
Power read by US board memory other
than set value.
Repeat cycle, if the problem persists
contact Technical Service Department
E901
Power detected by US board other than
set value.
Repeat cycle, if the problem persists
contact Technical Service Department
E957
UPDATING US IN PROGRESS
E999
Voluntary manual interruption of a
programme using 'STOP' button.
Содержание Tethys H10
Страница 1: ...OPERATOR S MANUAL...
Страница 37: ...EN OPERATOR S MANUAL 37 SERVICE This menu is intended for the technical service department...