Display the Ethernet configuration page:
Make sure that Automatic DHCP
configuration is selected.
With this selection all the numeric fields present on the screen are disabled (they get grey).
With this setting at each start the machine requires the network DHCP server its own
configuration using the DHCP protocol.
According to DHCP server configuration the numbering received may change at each start.
The TCP-IP number assigned to the machine appears on the upper blue bar in the first
screen (Home).
Type in the number present on the device followed by "." and by the TPC port number
(default 10080) in the web browser address toolbar of your PC i.e.: 10.20.8.115:10080:
The programme will display the web page:
(in the example the URL is http://10.20.8.115:10080)
It is usually possible to set DHCP server in order to always assign the same IP number to a
given device or to assign the same number to a given device for a predetermined period of
time. For these settings refer to the instruction manuals of your DHCP Server or of the local
network Internet router.
For these settings it is necessary to know the 'MAC Address' of the thermo-disinfector that
appears in the lower left corner of the Ethernet setting screen.