EN
OPERATOR'S MANUAL
57
2) CONNECTION TO A LOCAL NETWORK EQUIPPED WITH DHCP SERVER, WITH
THERMO-DISINFECTOR CONFIGURED WITH STATIC IP.
In order to avoid checking often the TCP-IP number assigned dynamically from a DHCP
Server, it is possible to assign manually a fixed number of the dynamic numbering of the local
network.
To avoid conflicts it is essential to:
- configure DHCP Server so that it does not assign the selected number to other devices.
Or
- assign statically a number out of the range assigned by the DHCP server to the thermal-
disinfector.
For the information needed for a correct configuration check DHCP server settings of the
local network.
To assign statically an IP address to the thermal-disinfector:
Access 'Data management' menu.
Display the Ethernet configuration page
Make sure that Automatic DHCP configuration is selected.
With this selection all the numeric fields on the screen are disabled (they get grey).
Take the first three numbers of the local network numbering, in the example above the first
three numbers are:
192.168.0
NOTE
AS AN ALTERNATIVE, IN WINDOWS SYSTEMS IT IS POSSIBLE TO USE
IPCONFIG COMMAND FROM A 'COMMAND PROMPT' WINDOW
(ACCESSIBLE FROM PROGRAMMES -> ACCESSORIES) TO DETECT
THE LOCAL NETWORK CONFIGURATION.
Now it is necessary to set statically the new number as follows:
1) Select manual configuration
.
2)
Set the first three fields of the address with the values detected (e.g.: 192.168.0).
3) Assign the chosen number to the last value, e.g. 222 (out of the range assigned
automatically, avoiding 0 and 255).
4) Check that Subnet Mask field is set to 255.255.255.0.
5) Gateway address is not important for communications inside the network (set 0.0.0.0).
Then the complete IP address (in this example) is: 192.168.0.100.
To connect to the thermal-disinfector enter the new number in the web browser address
toolbar as shown above (http://192.168.0.100).
The programme will display the thermal-disinfector web page.
Содержание Tethys H10
Страница 1: ...OPERATOR S MANUAL...
Страница 37: ...EN OPERATOR S MANUAL 37 SERVICE This menu is intended for the technical service department...
Страница 39: ...EN OPERATOR S MANUAL 39...