14
OPERATOR'S MANUAL
EN
ELECTRICAL
CONNECTIONS
The thermo-disinfector must be connected to a socket of the electric system having adequate
capacity for the absorption of the device and properly earthed, in accordance with laws
and/or regulations in force.
The socket must be properly protected through magneto-thermal and differential circuit
breakers having the following characteristics:
Rated current I
n
230-240V/50Hz
10 A
220V/60Hz
10 A
120V/60Hz
15 A
Residual current I
Dn
230-240V/50Hz
0.03 A
220V/60Hz
120V/60Hz
WARNING
THE MANUFACTURER IS NOT RESPONSIBLE FOR DAMAGES CAUSED
BY THE INSTALLATION OF THE THERMO-DISINFECTOR WITH
UNSUITED ELECTRIC SYSTEMS AND/OR NOT PROPERLY EARTHED.
NOTE
ALWAYS CONNECT THE POWER CORD DIRECTLY TO THE POWER
OUTLET.
DO NOT USE EXTENSIONS, ADAPTERS OR OTHER ACCESSORIES.
WARNING
IT IS ESSENTIAL THAT THE ELECTRIC SYSTEM TO WHICH THE
MACHINE IS CONNECTED COMPLIES WITH THE REGULATION IN
FORCE.
WARNING
ANY ELECTRIC CHECKS AND PREPARATION OF SYSTEMS MUST BE
WORKMANLIKE PERFORMED BY SKILLED PERSONNEL, HAVING
PROVEN EXPERTISE AND AUTHORIZED TO OPERATE ON ELECTRIC
SYSTEMS.
COMPETENT STAFF SHALL CHECK THAT EARTH CONNECTION IS
EFFICIENT.