EN
OPERATOR'S MANUAL
15
HYDRAULIC SET-UP
WATER FILLING
WATER INLET CONNECTION
The device is equipped with mains water filling pipe.
Pipe is pre-set for tap connection, with integrated "Aquastop" system serving as a protection
against any water leak, with 3/4" gas threaded connector.
NOTE
TO AVOID BACK-SIPHONAGE TO THE WATER MAINS, INSTALL A NON-
RETURN DEVICE ACCORDING TO PREVAILING STANDARDS.
NOTE
IF FILLING PIPE MUST CROSS A TOP, ARRANGE A D=75 mm HOLE.
MINIMUM HOLE D=65 mm REQUIRES THE REMOVAL OF AQUASTOP
SOLENOID VALVE SHELL (A).
POSITIONING OF FILLING TAPS
Filling taps must be positioned near the device, in a position accessible to the user.
For the positioning of filling taps there are no special constraints of height, except the length
of the supplied pipes, approx. 2 metres.
NOTE
TO PREVENT THE RISK OF CLOGGING OR DAMAGE, IF WATER
PIPELINE IS NEW OR HAS BEEN INACTIVE FOR A LONG PERIOD OF
TIME, MAKE SURE THAT WATER IS CLEAR AND WITHOUT
CONTAMINATION BEFORE CARRYING OUT WATER CONNECTION
HYDRAULIC SYSTEM
REQUIREMENTS
Check that the water supply pressure is within the required limits:
- minimum pressure 2 bar
- maximum pressure 5 bar.
NOTE
TO AVOID ANY FLOODINGS DUE TO POSSIBLE FAILURES, WE
RECOMMEND TO CLOSE THE MAINS WATER TAP WHEN THE DEVICE
IS NOT OPERATING, IN PARTICULAR AT NIGHT.