76
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET PROCEDURE
ALARMS (category A)
A022
door lock microswitch failure
(OFF-OFF)
dooR loCk pRoblem
1
A032
problem with the level sensor
h2o leVel pRoblem
1
A101
Temperature sensor pT1 failed
(
sterilization chamber
)
PTx INTERRUPTED
1
A102
Temperature sensor pT2 failed
(
steam generator
)
PTx INTERRUPTED
1
A103
Temperature sensor pT3 failed
(
heating element
)
PTx INTERRUPTED
1
A105
Temperature sensor pT5 failed
(
conductivity measurement compensation
)
PTx INTERRUPTED
1
A111
Temperature sensor pT1 short-circuited
(
sterilization chamber
)
PTx SHORT-CIRCUITED
1
A112
Temperature sensor pT2 short-circuited
(
steam generator
)
PTx SHORT-CIRCUITED
1
A113
Temperature sensor pT3 short-circuited
(
heating element
)
PTx SHORT-CIRCUITED
1
A115
Temperature sensor pT5 short-circuited
(
conductivity measurement compensation
)
PTx SHORT-CIRCUITED
1
A116
adC error
adC error
1
A120
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A121
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A122
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A123
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A124
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A125
Reference heating element acquisition chain
fault
ReFeRenCe heaTInG elemenT
FaulT
1
A201
Pre-heating not executed within timeout (ste
-
am generator)
GeneRaToR WaRm-up pRoblem
2
A202
Pre-heating not executed within timeout (tube
bundle heating element)
heaTInG elemenT WaRm-up
pRoblem
2
A203
pre-heating slow (steam generator)
GeneRaToR WaRm-up SloW
2
A204
pre-heating slow (tube bundle heating ele-
ment)
heaTInG elemenT WaRm-up
SloW
2
1
= ok (warning)
2
= ok + door unl ReSeT