12
– Remove the cap holding clip on the rear of the autoclave.
– Fit the plastic tube on the elbow union (provided).
–
Fit the union and then refit the clip.
– Fasten the clip provided to the drain siphon.
– Cut the hose to the right length and insert its free end into the centralized draining connector
locking it with the dedicated ring nut.
NOTE
make SuRe The Tube IS noT benT, CRuShed oR obSTRuCTed In any
Way.
The following diagram provides an indicative arrangement of the components:
NOTE
The ConneCTIon poInT To The CenTRal dRaIn muST be loWeR
Than The STeRIlIzeR'S SuppoRT SuRFaCe. oTheRWISe, The Tank
may noT empTy CoRReCTly.
NOTE
FoR auTomaTIC FIllInG SySTemS (pump, puRe 100, puRe 500, eV7)
The dIReCT dRaIn muST alWayS be ConneCTed.
ThIS aCTS aS an oVeRFloW, and, IF FoR Some ReaSon, The FIllInG
SySTem doeS noT STop, The WaTeR SeepS FRom The Feed Tank
InTo The dRaIn Tank and FloWS ouT The dIReCT dRaIn, ThuS pRe-
VenTInG FloodInG.
DIRECT CONNECTION
TO A CENTRALIzED
DRAINING POINT
Support
plane
To the centralized
draining point
drain
siphon
Clamp