Novodiag® System User Manual (NVD-SYST-UM)

 

 

V4-0 – April 2020 

29

 

Once  the  run  is  completed,  the  circle  indicates  the  completion  by  a  green  colour  and  a  symbol  indicates  the 
outcome  of  the  run  (see  Table  7  for  colour  and  symbol  coding).  Result  interpretation  is  detailed  in  each  Test 
specific Instructions for Use. 

6.5.

 

Shutting down the Novodiag® System 

The Touchscreen Computer can be shut down by using the close button located in the top right corner of the 
Software’s login screen. If you are logged in to the Novodiag® Software, you must first log out to access the close 
button. The dialog prompts to confirm the shutdown of the computer. Tap “OK”. The Novodiag® Instrument can 
be shut down by switching off the power switch behind the instrument. 

7.

 

Preventive maintenance and controls 

7.1.

 

Preventive maintenance 

No general maintenance is required for the Novodiag® System other than periodic cleaning: 

 

Clean the Novodiag® equipment according to the Chapter 8.1. 

 

Periodically  (once  a  month)  check  and  blow  dry  air  onto  the  ventilation  system  on  the  sides  of  the 
Novodiag® Instruments. 

CAUTION 

Before  cleaning  the  Touchscreen  Computer,  close  the  Novodiag®  Software  application  to  avoid 

accidental opening of the Novodiag® Instrument. 

7.2.

 

Controls 

7.2.1.

 

Internal Control 

Each Novodiag® Test Cartridge includes an internal process control, which acts as a control through the whole 
analysis. The process control contains synthetic gene target and is applied in dried format on the sample filter in 
the  cartridge  during  manufacturing.  For  this  reason,  the  process  control  is  subjected  to  the  same  sample 
preparation process as the sample.  
 
The process control detection will fail if inhibition or other disturbing factors affect PCR amplification. For the test 
to be successful, either process control, target(s) or both need to give a positive result, otherwise the test result is 
invalid. 

7.2.2.

 

External Controls 

To ensure the proper functioning of the Novodiag® System, it is highly recommended to periodically run external 
controls according to the instructions provided in Chapter 6.

 

NOTE External controls are not provided by Mobidiag. 
 
It is recommended to use positive controls at regular intervals and in accordance with laboratory guidelines. They 
can also be used in case of a suspected malfunction of an instrument or cartridge batch. The user should prepare 
specific positive controls for each test according to laboratory guidelines. Positive controls should always give a 
positive  result.  A  negative  result  from  a  positive  control  indicates  a  possible  malfunction  of  the  instrument  or 
defective cartridge. In this case, please contact Mobidiag representative. 
 
As  a  negative  control,  appropriate  sample  buffer  may  be  used  as  the  sample  (refer  to  Novodiag®  Test  specific 
Instructions  for  Use  for  more  information).  A  negative  control  is  used  after  installing  the  instrument  or  after 
maintenance carried out by a Mobidiag representative, to ensure the proper working of the instrument. It is also 
recommended for use at regular intervals, according to laboratory guidelines. A negative control can also be used 
in case of suspected nucleic acid or sample contamination in the working area or equipment used.  

Negative  controls  should  always  give  a  negative  result.  A  positive  result  from  a  negative  control  indicates  a 
possible  nucleic  acid  or  sample  contamination  in  the  working  area  or  equipment  used.  In  this  case  clean  the 
working area and used equipment with nucleic acid decontaminants and disinfectants, according to laboratory 
guidelines  and  procedures  described  in  Chapter  8.1.  In  case  of  an  invalid  result  from  a  negative  control,  the 

Содержание Novodiag NVD-BCS-A NVD-TSCA-A

Страница 1: ...Novodiag System NVD INST B NVD DSW For in vitro diagnostic use For use with Novodiag Tests User Manual English Version 4 0 Issued in April 2020...

Страница 2: ......

Страница 3: ...xtraction 7 3 3 Amplification 7 3 4 Microarray hybridisation 7 3 5 Detection and analysis 7 4 Configuration and setup 7 4 1 Environmental requirements 7 4 2 Installation requirements 8 4 3 Packaging c...

Страница 4: ...rument maintenance 31 8 4 Disposal 32 9 Precautions limitations and safety hazards 32 9 1 Precautions and limitations 32 9 2 Safety Information and warnings 32 9 3 Electromagnetic compatibility 33 9 4...

Страница 5: ...denotes a hazard It calls attention to an operating procedure or practice which if not performed correctly could damage the instrument or invalidate results NOTE A note provides important helpful or...

Страница 6: ...f symbols These symbols can be found on the Novodiag System its associated components or throughout this manual Use the definitions below to interpret the symbols see Table 2 Table 2 Definitions for s...

Страница 7: ...and microarray hybridisation to detect and identify multiple nucleic acid targets from clinical samples providing qualitative result which is then displayed on the Novodiag Software The device s func...

Страница 8: ...ht below each slot allows to monitor the Novodiag Instrument slots status see Chapter 4 5 1 for light status correspondence Up to four Novodiag Instruments can be stacked and controlled by a single To...

Страница 9: ...ed users only For further information about Novodiag Software see Chapter 5 2 4 4 Additional components Barcode Scanner ref NVD BCS A Arm ref NVD TSCA A Post ref NVD TSCP A fixtures and strip The Barc...

Страница 10: ...time PCR and microarray detection Each one amplifies and detects a set of targets or is a duplicate for targeted test 3 Principle of operation Novodiag System uses a combination of real time PCR and...

Страница 11: ...red based on a proprietary optical module 3 4 Microarray hybridisation For Novodiag Test panels the amplified material is transferred to the microarray for hybridisation Amplicons bind to a set of com...

Страница 12: ...x 50 x 50cm stack 40 x 20 x 30cm computer 45kg 600x2 65 computer 2 1 Two instruments Arm 125 x 50 x 50cm 50kg 600x2 65 computer 2 1 Three instruments Arm 125 x 50 x 75cm 70kg 600x3 65 computer 2 1 Fou...

Страница 13: ...yellow cartridges Power cables and connection cables The Touchscreen Computer the Barcode Scanner and the Post or Arm are delivered in separate packages CAUTION The Novodiag Instrument unpacking must...

Страница 14: ...the whole stack and disconnect it from the power supply 4 5 2 Adjusting the Touchscreen Computer on Post or Arm You can adjust the orientation of the Touchscreen Computer on the Post or on the Arm by...

Страница 15: ...nts Do not change screensaver settings 4 5 4 Starting up the Novodiag Software To start up the Novodiag Software double tap the Novodiag Software icon on the desktop The Novodiag Software can be start...

Страница 16: ...min user are displayed in Table 5 See chapter 5 6 2 user management for details on how to create users and modify their permissions Mobidiag retains one admin user account for service personnel to use...

Страница 17: ...m Also before starting a run please check the username on the screen to make sure that you are the right user logged in The Home View reflects the physical setup of the Novodiag Instrument and is divi...

Страница 18: ...iew Instrument in stack not displayed on instrument view Instrument not connected Instrument disabled Test creation Slot empty ready for run Test creation Slot selected next run will be created to thi...

Страница 19: ...by the symbols inside the circle as explained in Table 7 Table 7 Instrument view symbols Slot view Status Slot empty Slot initialising Run in progress 31min remaining Slot disabled Slot view Status Ru...

Страница 20: ...w tap the View test results button on the bottom of the slot pane see Figure 14 This brings you to the Results view Alternatively result history is available by tapping the Results button at the top o...

Страница 21: ...Sample ID Test type Date time and user Main Result POSITIVE NEGATIVE INVALID FAILED List of targets detected No targets detected Remarks if related to the targets Figure 15 Result view By tapping the...

Страница 22: ...selected run see Figure 15 Runs can also be filtered by tapping the Edit filters button on the Filter pane see Figure 17 The filters are Test type drop down menu Result type drop down menu Operator d...

Страница 23: ...file Mobidiag representative may request you to deliver log files for troubleshooting failed runs see Chapter 10 Figure 19 Result View Exporting options As a default the exported files csv PDF or log...

Страница 24: ...ew user tap the User Management button Create a new user by tapping the Add new user button on the bottom of the user list Add username pin code and credentials The credential details are indicated in...

Страница 25: ...he results The time for logout can be adjusted by tapping the wanted logout time see Figure 22 Figure 22 System Setup view Settings tab As a default setting the Novodiag Software saves all exported fi...

Страница 26: ...use by tapping ENABLED If a default printer is set the result is sent to the printer in PDF format Contact Mobidiag representative for setting up the default printer 5 6 4 Slot Management You can ena...

Страница 27: ...rformed Once installed test plugins can be enabled and disabled from the Test Management view by selecting the test and tapping the ADD TEST or REMOVE button see Figure 26 Enabled test plugins go from...

Страница 28: ...n the Novodiag Instrument by switching on the power switch behind the instrument Turn on the Touchscreen Computer by using the button on the bottom right of the screen Login to Windows occurs automati...

Страница 29: ...Test Cartridge Do not place any stickers or labels directly on the cartridge Do not write outside the cartridge label and do not impede barcode reading see Figure 4 6 4 Starting a run Bring the Novod...

Страница 30: ...g the code and tapping Enter NOTE The Novodiag Software recognises automatically the Novodiag Test Cartridge barcode and uses the relevant Test plugin to apply the right protocol The Novodiag Software...

Страница 31: ...es the information about previously used slot and offers the next available slot for the run NOTE The Novodiag Instrument detects the cartridge presence and ejects the tray if the cartridge has not be...

Страница 32: ...e 33 NOTE The estimation of the remaining time is based on the run history for a given test In case the type of test is run for the first time the remaining time displayed may be inaccurate Figure 33...

Страница 33: ...R amplification For the test to be successful either process control target s or both need to give a positive result otherwise the test result is invalid 7 2 2 External Controls To ensure the proper f...

Страница 34: ...false positive by the Novodiag System WARNING If the cartridge contains potentially infectious material the risk of biohazard contamination exists in addition to DNA RNA contamination 8 1 1 Cleaning M...

Страница 35: ...contamination clean the bench tops other used equipment and working areas as follows 1 Wear appropriate PPE such as a laboratory coat and gloves 2 Wet a paper towel with DNA RNA degrading agent and w...

Страница 36: ...out an approval from Mobidiag As detection areas on the Novodiag Test Cartridge are sensitive to damage and smear do not touch or get in contact with sharps or dirty material While a run is in progres...

Страница 37: ...rds WARNING Breakable glass slide microarray is present inside some of the Novodiag Test Cartridges refer to the Novodiag Test specific Instructions for Use It can expose you to sharp hazards and caus...

Страница 38: ...aged during transport Discard cartridge and use a new one Protective cover not fully removed or white membrane damaged Discard cartridge and use a new one White blocker not properly secured before run...

Страница 39: ...Retry to read the barcode If the problem persists type the barcode manually If the problem persists contact Mobidiag representative Not enough disk space Remove any large files such as run reports fr...

Страница 40: ...ed Screen is black Wake up the Touchscreen from sleep mode by tapping the screen Touchscreen not responsive Force a shut down and restart the Touchscreen Press and hold the power button until the scre...

Страница 41: ...6 1000 pixels camera Temperature control Thermocycling devices Operating from 15 110 C Ramp rates heating 7 C s Cooling 5 C s Environmental specifications Indoor only Temperature 18 C 30 C Humidity 15...

Страница 42: ...ratio Brightness 220 nits Resolution 1366 x 768 Touchscreen Resistive touch PCAP Storage device HDD 1x 2 5 SATA 500Gt External I O ports USB 4x USB3 0 2x USB2 0 LAN 2x RJ45 Power Power adapter 65 W 1...

Страница 43: ...ks or registered trademarks of Mobidiag Oy Other trademarks mentioned in the User Manual or Instructions for Use are the property of their respective owners No rights to use any of the aforementioned...

Страница 44: ...1 4 0 Ch 5 5 1 Log file exporting is available for results that are maximum of one year old only Ch 5 6 3 Addition of an automatic printing feature Ch 6 3 Defining of warnings and precautions related...

Страница 45: ...catalogue number NVD INST B NVD DSW NVD TSC A NVD BCS A NVD TSCA A NVD TSCP A Go to our website www mobidiag com For more information Mobidiag Ltd Keilaranta 16A FI 02150 Espoo FINLAND www mobidiag c...

Отзывы: