![Mobidiag Novodiag NVD-BCS-A NVD-TSCA-A Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/mobidiag/novodiag-nvd-bcs-a-nvd-tsca-a/novodiag-nvd-bcs-a-nvd-tsca-a_user-manual_1815166012.webp)
Novodiag® System User Manual (NVD-SYST-UM)
V4-0 – April 2020
8
4.1.2.
For Novodiag® Test Cartridges
Refer to the Novodiag® Test specific Instructions for Use.
4.2.
Installation requirements
Table 3. Requirements for each installation configuration
Space requirement
(W x D x H)
Weight
requirement
Max power
consumption (W)
Number of
power sockets
LAN
(optional)
Single instrument
+ Post
100 x 50 x 25cm
(instrument)
+ 40 x 20 x 30cm
(computer)*
25 kg
600 + 65
(computer)
2
1
Two instruments
+ Post
100 x 50 x 50cm
(stack)
+ 40 x 20 x 30cm
(computer)*
45kg
600x2 + 65
(computer)
2
1
Two instruments
+ Arm
125 x 50 x 50cm**
50kg
600x2 + 65
(computer)
2
1
Three
instruments
+ Arm
125 x 50 x 75cm**
70kg
600x3 + 65
(computer)
2
1
Four instruments
+ Arm
125 x 50 x 100cm**
90kg
600x4 + 65
(computer)
2
1
*Touchscreen Computer on Post can be placed directly on top of Novodiag® Instruments or beside it.
**The Arm configuration holds the Touchscreen Computer above the workspace.
As depicted in Figure 6 and Figure 7, a minimal space must be free around the Novodiag® Instrument to allow air
circulation as well as provide some room for the Arm in case of stack configuration.
Figure 6. Top view of the Novodiag® Instrument