1 General notes
The unit CLIMACELL should be used in applications, in which changing the temperature in a
sample in range from 0
°
C to 99.9
°
C and the relative humidity (RH) from 10 % to 95 % in
various time modes is needed.
CLIMACELL may be used in several ways, so please, read this manual carefully.
MMM’s cooling incubators CLIMACELL are equipped with a microprocessor control with the
following features:
1. Fuzzy-logic
temperature controller and a thermo-sensor PT 100.
2. Humidity control by PID regulator with
fuzzy logic
and
a humidity capacity sensor
.
3. 4 preset programs P1 to P4, 2 programs P5 and P6 to set containing up to 16 steps.
4. Electronic timer with different time modes, see description of programs (page 10).
5. Adjustable fan speed from 0 to 100 % in 10 %-steps.
6. Printing the protocol with real temperature and time values via interface RS 232 C.
7. Chip card reader to store data P1 to P4, P5, P6 and to record to the memory .
8. Foil covered touch buttons for easy programming.
9. 32 digit LCD display, illuminated. All parameters seen at once. (To your information, the
display is already set up to be used in product ranges coming out in future.)
CLIMACELL is manufactured of high grade materials with the latest technology and is
subjected to strict final tests leaving the plant. The inner chamber housing as well as the
shelves are made of stainless steel. The coating is water based, gray on top of galvanized
zinc coated sheet metal, safe for the environment.
2 Important instructions
2.1
Unpacking and check
After unpacking the unit, please, check, whether the unit and its facilities are complete and
undamaged.
An eventual damage must be immediately reported to the transporting company.
During the manipulation – in case of lifting the cabinet etc. – the cabinet cannot be hold by
the rail or door. The cabinets of volumes 404 and 707 should be lifted by means of delivered
hooks, the rolls are designed for local moving, not for longer transport.
The standard delivery consists of the temperature cabinet,
two sieves,
one chip card SO for setting the safety thermostat,
three chip cards for storing the set programs
hose for connecting the input water
(order No.: 0671574),
hose for outlet of the waste water
(order No.: S461200),
dish for collecting the condensate