10.7 Potential-free contact for alarm reports
It's outlet is placed in the connector in the rear foot, a voltage up to 24 V/ 1A can be
connected to it.
10.8 The inner switched socket
The socket is placed inside the chamber on the side wall. Details of the switching control see
par. Basic setting of the units Climacell – Service 06, connection and other details are given
in the following diagram and text.
Connecting the connector to the built-in device.
Warning:
The plug must be connected only by a person with a sufficient qualification.
When the plug is plugged in and out of the socket, or when manipulated with the inner
equipment, CLIMACELL must be switched off (the mains switch in the position-„0“).
The socket-plug contact must be tightened in order to prevent disfunctioning of the protective
system.
After disconnecting the socket-plug contact fit carefully the protection covers on the socket
and the plug so that impurities and humidity would not get inside.
10.9 Exposition lighting in the door
There is an exposure fluorescent lighting placed in the door, whose intensity is controlled
from 0 % to 100 % in 10 %-steps, details see chapter Basic setting, point Setting the lighting
– type with door lighting.
PLUG
TO DEVICE