Operating instructions CLIMACELL Page 36
V 1.07_Blue Line CLIMACELL_PCH_en_np_mmm
!
0208
10 Optional
equipment
10.1 Inner
lighting
Inner space is illuminated with thermo-resistant bulbs - their switch is placed on the door.
10.2 Bushings of diameter 25, 50, 100 mm
The bushings are normally placed cca in the middle of the side (right or left) wall of the
chamber. The bushes are metallic, closed with a special plastic plug from the outer side,
which enables passing through of wires etc. from the outer space to the chamber.
Recommendation: the used unit should be provided with a bushing with corresponding
dimensions, if user wishes to measure temperature or RH inside the chamber by means of
sensors, that are connected with an independent measuring device by means of wires; the
user pulls the wires through the bushing.
10.3 Lockable
door
The lock is placed on the upper part of door's surface near the closing mechanism.
10.4 Left
door
This is the symmetric mirror version of the right door. The cabinet of 404 liter is not delivered
in this version.
10.5 Independent sensor PT100
Independent flexible sensor is used for measuring of material temperature directly in the
chamber. The date of temperature on this sensor (PT2) is displayed on LCD in the position of
time date, while these two data are altering.
The corresponding printed protocol has two lines, first line - beginning with digit 1 - shows the
temperature on the sensor of regulator (PT1), second line beginning with digit 2 shows the
temperature on the sensor PT2.
In case of failure of the contact between PT2 and regulator the display shows Error 01.
10.6 Communication SW DS Com for PC under Windows
The software DS Com is designed to record the temperature course in the thermal cabinets.
Data obtained during the regulation are displayed in form of a graph (with time on the
horizontal axis and obtained data on the vertical axis). The software enables to follow the
regulation in real time, to store the regulation course to a file on the disc and to view the
already stored files.
Operation instructions of this software are delivered together with the installation software,
minimal requirements on PC hardware for the program DS Com 2.1 are:
•
IBM-compatible PC (386 and above)
•
Operation system Windows 95 and above
•
80 MB RAM
•
one-hour-record of data requires free space of about 60 KB on the hard disc
•
free serial port