Operating instructions
CLIMACELL
Page 27
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
If the cycle selection is active on LCD at the moment of the start, see fig. Display - state of
the program selection - E: cXX, the program comes again to the start after the last segment
is over and the whole program is repeated. The definite number of program cycles
corresponds with the number behind the letter “c“. The segment 0 can be counted into the
cycle, it can be skipped as well.
The time is - as distinct from P5 - not defined as the length of the segment, but it is defined
by the hour, minute, day and month of its end, that means absolutely with respect to the real
time. If the time and the date of the segment is set lower than the time and date of the
previous segment, the segment will be finished in the next year. Set the relevant time and
date including the year in the service mode.
The rise to the set RH starts after reaching the
set temperature.
Range of parameters:
time
t0 to t15 each of segments up to a year
ramp
r1 to r15 0 to 3.2 °C/min for heating
-1.0 to 0 °C for cooling
temperature
T1 to T15
0,0 to 99,9 °C
fan speed
V1 to V15
0 to100 % with 10% steps
relative humidity
RH1 to RH15
10 % to 90 % with 1 %
light
L1 to L15
ON/OFF (only for the type CLC-L)
number of cycles C
2 to 99 (eventually
∞
- which is the next
position after 99)
Display - state of the program selection
A B C D E
F G H I
A: P6
name of the program
B: XX
number of the set segment
C: X
type of the set segment:
x segment will be skipped over
> not the last, the end length - see E, I
>I the last, the end length - see E, I
∞
the last, infinite length
D: XX . X
the required temperature of the set segment (! the end temp.)
E: tXX : XX the hour and minute of the set segment end, deciding is the ramp or
TXX : XX the hour and minute of the set segment end, deciding is the time
r
±
X . XX the ramp = required speed of the set segment, deciding is the time or
R
±
X .XX the ramp = required speed of the set segment, deciding is the ramp or
cXX
the number of cycles of the whole program
F: XXX
the fan speed of the set segment
G: XX
disengaged
H: XXX
the light of the set segment
I: XX . XX
month and day of the set segment end