Operating instructions
CLIMACELL
Page 13
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
4.4.9 Service 08 – Selecting the water quality
Enter the service by pressing START/STOP. On the upper line of the display: „Water
Quality“
On the lower line switch over by pressing
▲
: either „Tap Water“ or „Destilled Contai“.
Conform the setting by pressing START/STOP and quit the service 08.
4.4.10 Service 09 – Canceling the service regime
Cancel the service regime by pressing START /STOP. Now you can select and start the
programs.
4.4.11 Setting the lighting – type with door lighting
Programs
P1 to P4, P5, P6
– type with dimmer
Concerning extension of following chapters:
Setting the program P1
Setting the program P2
Setting the program P3
Setting the program P4 – segment T1
Setting the program P4 – segment T2
Description of the program P5 and its control
Description of the program P6 and its control
By pressing
◄►
place the cursor to the position
☼
and by pressing
▲
select the lighting.
P1 to P4
: after reaching the set temperature in the just adjusted horizontal segment – or at
P3 also at the time delayed period – there is possible to set the illumunation at steps of 10%
of maximum illumination. 0% (totally off) is at display shown as „OFF“, 100% is shown as
„ON“. The illumination is constant during the whole time as the segment continues.
P5, P6
: there is possible to set the illumination at steps of 10 % of maximum illumination at
every (horizontal or cross) segment. 0% (totally off) is at display shown as „OFF“, 100 % is
shown as „ON“. The illumination is constant during the whole time as the segment continues.
The reference value of illumination (when 100 % and 22 °C is set) on the level of the entry
into the chamber (in the middle of the entry area) – measured with the luxmeter sensor
perpendicularly towards the impinging rays, white day light: CLC 111: 6000 lx
±
5 %, CLC
222: 10 000 lx
±
5 %, CLC 404 and 707: 13 000 lx
±
5 %
4.4.12 Setting the relative humidity in programs P1 – P4
During the setting of temperature and time in the given horizontal segment set the required
value of relative humidity in 1 %-steps in the range of RH between 10 % and 95 % at the
position of RH on the lower line of the display – with limitations given in diagram in
paragraph
Setting the operating conditions of temperature and relative humidity and
their limitations.
Rise to the set RH starts after reaching the set temperature.
Notice:
Switch off the RH regulation during the run of the program with the already set RH value
according to the following example: If the cabinet operates e.g. at 20 °C and 86 % RH,