Operating instructions
CLIMACELL
Page 31
CLIMACELL_PCH_en_np_mmm
!
0208
V1.07_Blue Line
4.11 Printing the protocol
Printing the protocol with printer CITIZEN, model iDP 3110-24 RF-A.
By setting the printer interval you give the time intervals of printing actual values of
temperature in the chamber. Length of this interval is displayed in the head of the protocol.
Printer interval can be set - see paragraph Basic Setting, point Service 04.
1. Connect the printer to the thermal cabinet by means of the cable with serial connector
(RS-232C) into the socket for the printer (see chapter Description of the cooling
incubator, par. Connector RS-232C for connecting the printer), the other end of the cable
connect to the connector in the rear part of the printer. Connect the printer to the mains by
means of the adapter.
Connect the printer to the mains by means of an adapter. Switch
the switch on the right side of the printer on. The control light
POWER
and
SEL
switches
on. The printer is in mode ON-LINE. You can shift the paper by pressing the button LF
(can be carried out only in the mode OFF-LINE, i.e. after pressing the button SEL, the
control light
SEL
switches off. To renew the mode ON-LINE the button SEL must be
pressed, the control light
SEL
switches on). The printer prints only in mode ON-LINE!
2. Set required interval downwards or upwards according to point Service 04 in chapter
Basic setting.
3. There is a head printed by the printer, the head comprises: type of unit, set temperature,
eventually set ramp, number of cycles, ventilator rate, RH, light and the selected printing
interval. Below the head following values are printed:
time since the start of the program, real temperature in the chamber (below eventually
temperature on the flexible sensor), RH.
A new head is printed if:
- the cooling incubator is started
- some parameter has been changed
4. You can stop the printing by setting the printing interval to
Printer off.
5. In case of a mains fall-out and restoring the supply again there is
→
Mains fall-out
printed. The printing intervals start to be counted since the moment of restoring the
current supply.
6. In case of fall-out of the printer or switching the printer off there is no report printed after
the switching on or restoring the supply.
Setting the DIP switches of the printer Citizen iDP 3110-24 RF-A:
1., 3. and 4. DIP switch (from the left) – pin is in the lower position
2. DIP switch (from the left)
– pin is in the upper position
4.12 Checking the function of the safety thermostats
a) safety thermostat of the chamber
According to the procedure described in Service 07 – Setting the safety thermostat, set the
limit temperature 85°C and confirm with the chip card SO. In P1 set the temperature 80°C
and start the program. After reaching the set temperature wait ca. 10 minutes, stop the
program and enter the Service 07; in Service 07 set the limit temperature to 70°C, confirm
the setting with the chip card SO. The unit operates according to the description in Service
07 -Setting the safety thermostat: on the display “Safety thermost” etc.
b) safety thermostat of the steam generator
Disconnect Climacell from mains, dismount the fixing screw of the steam generator and shift
the generator forwards. Unscrew completely the nut fixing the safety thermostat sensor and
take the sensor with the nut and the rubber sealing out of the generator. Disconnect both
wires from the contacts of the safety thermostat switch box and connect them to the
terminals of the gauge, which is set to measure resistance. Under normal conditions (T = 23
°C) the resistance = 0.