5.2 Electrical Connections
Insert the 15-way D-Type socket on the output lead of the power supply unit into
the 15-way D-Type connector labeled
“POWER”
on the rear panel of the HPQ2-IP.
Attach the line cord to the power supply. The power supply will automatically set
itself to operate with the local line voltage.
When power is applied, the LED on the rear of the HPQ2-IP will illuminate.
In the case of an Ethernet connection, connect a cat5 patch lead to the socket
marked
“ETHERNET”
. If a direct connection to the host PC is required, use a
crossover cable. If connecting via a network hub, a standard patch lead may be
used.
If required, connect your X-Trip signal cable to the socket marked
“X-TRIP”
.
Care should be taken in routing and securing all cables. Avoid running any of the
signal cables next to mains power cables or sources of electrical noise.
5.3 IP Address
The default IP address of the HPQ2-IP is 192.168.0.250.
Details on how to change this IP address can be found in the manual supplied with
your RGA operating software.
5.4 Re-setting the IP address
To re-set the unit back to its default IP address, carry out the following procedure:
1.
Push the “reset” button
2.
Immediately push and hold the “preset” button until the unit emits a
bleep. The IP address has now been re-set.
Further details on using and assigning an IP address can be accessed from the built-
in help files contained in the Process Eye/EasyView software, or in the
Troubleshooting section of this manual.