HPQ2-IP SP101009.102 – 10 October 2012
MKS Instruments, Spectra Products
17
3.7 Digital I/O Connector
The Digital I/O connector is a 25-way D-Type socket mounted on the rear panel of
the unit and labelled
DIGITAL I/O.
It is used to provide alarm output signals and
process trip signals. It can also be used to connect accessories such as a Remote
Vacuum Controller, analogue output module, analogue input module and Valve
Controller.
The Digital I/O port provides:
Two 8 bit bi-directional ports, one interrupt / strobe input
+5V fused at 100mA
±
15V fused at 100mA
DIGITAL I/O Connector Pin Assignment
Pin
Description
Notes
1
H1 interrupt / strobe
2
PA1
3
PA3
4
PA5
5
PA7
6
PB1
7
PB3
8
PB5
9
PB7
10
0V Digital
11
0V Analogue
12
-15V fused
13
+15V fused
14
PA0
15
PA2
16
PA4
17
PA6
18
PB0
19
PB2
20
PB4
21
PB6
22
+5V fused
23
Not Connected
24
Not Connected
25
Not Connected
Note that the total power consumption on each rail (+5V and
±
15V) for
both the Analog and Digital I/O ports must not exceed 100mA.