946 Controller Instruction Manual
pg. 32
CM
MFC
Scale Factor
Zero
FD
1.00
NO
FlowRate SP
NO
2.00E+2
Relay
Relay n
NO
YES
Enable
DIR
SET SP
Hyst
OFF
BELOW
1.00E+01
1.50E+01
0.1 to 10
Open
Message
Display 2 Relays, Ch# auto detected
ON
OFF
ABOVE
BELOW
Display warning message
1% to 95% of its full scale
Nominal Range
1.00E+3
0.1 to 1.00E+6
Range
1.00E+3
OP Mode
Setpoint
Display only
Close
Setpoint
PID Ctrl
RatioM
RatioA
0 to Nominal Range
Operation Voltage
+15 to +24V
ON
+15 to +24V
+/-15V
Channel
Setup
Range
Factory Default
Manual Zero
1000
NO
Auto Zero
NO
A1
Relay
Relay n
NO
YES
Enable
DIR
SET SP
Hyst
OFF
BELOW
1.00E+01
1.50E+01
0.02 to 20,000
NA
Ion gauge Channel
Message
Display 2 Relays, Ch# auto detected
ON
OFF
ABOVE
BELOW
Display warning message
1% to 95% of its full scale
Operation Voltage
+/- 15V
Figure 6-7 946 Channel Setup Setting Parameters, their Default Values and Ranges
6.5.2 Setup for a Capacitance Manometer
A capacitance manometer module will be automatically detected on power-up. At the same time, the
connection of the capacitance manometer to the module will be checked. If no capacitance manometer is
connected, - - - - will be displayed. A CM module operates up to two Capacitance Manometers, and the
total combined current requirement of operating sensor(s) must be 1 amp or less.
Refer to Figure 6-8 in setting up a capacitance manometer.
1. CM Type
ABS (absolute) can be selected.
2. Input Voltage
The Input Voltage for the controller is same as the maximum analog output voltage of the
capacitance manometer. To select, move the cursor to the Input Voltage box, press
Enter
, and to
highlight the parameter. Use either
or
to select the correct voltage. Press
Enter
to save the
setting, either 10 or 5 V.