946 Controller Instruction Manual
pg. 29
When selecting A from the front panel, only multiples of 1, 2, and 5 are allowed. Following are
some examples for A corresponding to P
max.
P
max
1×10
-3
2×10
-3
5×10
-3
1×10
-2
1×10
2
2×10
2
5×10
2
1×10
3
1.5×10
3
A
1E+4
5E+3
2E+3
1E+3
1E-1
5E-2
2E-2
1E-2
8.4E-3
If a 10V linear output is needed at a pressure different from that given above, it can
be entered via serial communication using the DLAn and DLT commands outlined in
Section. 9.14.
15. FV (firmware version) Information
The firmware version for all of the modules installed in the 946 is displayed when Display
is selected. The serial numbers for all detected boards are also displayed, as shown in
Firmware Version and SN
Slot A
1102114509
CC
1.00
Slot B
1103104503
CM
1.00
Slot C
1105083309
PR
1.00
Analog IO
1101154102
AIO
1.00
Comm
1102104501
Com
1.00
Main
1106031428
Main
1.00
Detected
board
Firmware
Version
Serial
Number
Figure 6-6 System Firmware and Serial Number Information Displayed
16. Display format
Display format allows selection of the front panel display format. There are three options:
Default, PatchZ, and HighR.
Default format: only significant digits are displayed.
PatchZ format: Zero(s) will be patched to enable 4-digit display for CM, and 3 digit
display for CC or HC and Pirani or Convection Pirani.
HighR format will display extra digit for CC and HC which may assist to identify pressure
changes.