Chapter 3
40
Micro-Ion
®
Module Instruction Manual - 356007-GP
3.12
RS-485 Operation and
Commands
This following sections of the Operation chapter provide detailed
information to use the Series 356 Micro-Ion Plus Module with the RS-485
digital interface. Topics covered are:
RS-485 Physical Layer (Section 3.13)
RS-485 Initial Setup (Section 3.14)
RS-485 Data Timing and Response (Section 3.15)
RS-485 Command Structure (Section 3.16)
RS-485 Command Sets (Section 3.17)
RS-485 Commands (Section 3.18 and Section 3.19)
3.13
RS-485 Physical Layer
Table 3-8
RS-485 Physical Layer Specifications
Function
Description
Arrangement
2-wire half duplex
Address range
0 to 63
Default address
01
Method for setting module address
• Use address switches to set value of 0 to 15 for address
• Send SA command to set value of 10, 20, or 30 for address offset
Maximum cable length
• 4000 feet (1610 meters)
• A common ground wire should connect all network devices for
long cable runs
Maximum number of devices in network
31 devices
Default baud rate
19200 baud (19.2 kbaud)
Data bits
8 data bits
Stop bits and parity
1 stop bit, no parity
Содержание 356 Micro-Ion Plus Series
Страница 2: ......
Страница 24: ...24 Micro Ion Module Instruction Manual 356007 GP...
Страница 58: ...Chapter 3 58 Micro Ion Module Instruction Manual 356007 GP...
Страница 68: ...Appendix A 68 Micro Ion Module Instruction Manual 356007 GP Figure A 1 Automatic Filament Switching Flow Chart...
Страница 70: ...70 Micro Ion Module Instruction Manual 356007 GP Figure A 3 Manual Filament Switching Flow Chart...
Страница 74: ...Index 74 Micro Ion Module Instruction Manual 356007 GP...
Страница 75: ......