66
Micro-Ion
®
Module Instruction Manual - 356007-GP
4.4
Replacing the Gauge
Assembly
Do not plug in or unplug any connectors with power applied to the module.
Disconnect power from the module before replacing the Micro-Ion gauge.
1.
Turn OFF power and disconnect all electrical connectors to the
module.
2.
Remove the module from the vacuum system.
3.
Unscrew the four Phillips-head screws until they disengage from the
gauge. See Figure 4-1.
4.
While holding the vacuum connection flange, gently unplug the gauge
and end plate out of the module as shown in Figure 4-2. The gauge and
end plate are shown in Figure 4-3.
5.
Insert the new gauge into the module by gently inserting the gauge pins
into the socket on the circuit board. Look at the gauge pin arrangement
to be sure you do not have the replacement gauge and end plate
assembly improperly rotated.
6.
Tighten the four Phillips-head screws.
7.
Install the module on the vacuum system.
8.
Calibrate the Micro-Ion Module at atmosphere (see page 35).
9.
If Micro-Ion gauge does not turn ON at 20 mTorr (2.66 x 10
–2
mbar,
2.66 pascal) with decreasing pressure, calibrate the Micro-Ion Module
at vacuum (see page 36).
4.5
Returning a Micro-Ion
Module for Service
If the product must be returned to the factory for service, request a Return
Material Authorization (RMA) from Granville-Phillips. Do not return
products without first obtaining an RMA. In some cases a hazardous
materials disclosure form may be required. The MKS/Granville-Phillips
Customer Service Representative will advise you if the hazardous materials
document is required.
When returning products to Granville-Phillips, be sure to package the
products to prevent shipping damage. Shipping damage on returned
products as a result of inadequate packaging is the Buyer's responsibility.
For Customer Service / Technical Support:
MKS Pressure and Vacuum Measurement Solutions
MKS Instruments, Inc., Granville-Phillips
®
Division
6450 Dry Creek Parkway
Longmont, Colorado 80503 USA
Tel: 303-652-4400
Fax: 303-652-2844
Email: [email protected]
Содержание 356 Micro-Ion Plus Series
Страница 2: ......
Страница 24: ...24 Micro Ion Module Instruction Manual 356007 GP...
Страница 58: ...Chapter 3 58 Micro Ion Module Instruction Manual 356007 GP...
Страница 68: ...Appendix A 68 Micro Ion Module Instruction Manual 356007 GP Figure A 1 Automatic Filament Switching Flow Chart...
Страница 70: ...70 Micro Ion Module Instruction Manual 356007 GP Figure A 3 Manual Filament Switching Flow Chart...
Страница 74: ...Index 74 Micro Ion Module Instruction Manual 356007 GP...
Страница 75: ......