Micro-Ion
®
Module Instruction Manual - 356007-GP
67
Appendix A Filament Operation
Table A-1
Micro-Ion Filament Operation Modes
Mode
Description
Automatic
• Filament logic operates filament 1 each time the Micro-Ion gauge turns ON
• If filament 1 becomes inoperable, the Micro-Ion gauge switches to filament 2
Alternating (default for Micro-Ion gauge
with yttria-coated iridium filaments)
• Filament logic alternates between filament 1 and filament 2 each time the
Micro-Ion gauge turns ON
• If a filament becomes inoperable, the module switches to the operable filament
• Default mode for Micro-Ion gauge with yttria-coated iridium filaments
Manual (default for Micro-Ion gauge
with tungsten filaments)
• Filament logic operates filament 1 each time the Micro-Ion gauge turns ON
• If filament 1 becomes inoperable, press the gauge button to turn filament 2 ON
(gauge LED will flash red)
• After the gauge button has been pressed, filament 2 turns ON
• Default mode for Micro-Ion gauge with tungsten filaments
Table A-2
Micro-Ion Gauge, Gauge LED, and Filament States in Automatic or Alternating Mode
Micro-Ion Gauge State
Gauge LED State
Filament State
OFF automatically
Amber
Micro-Ion gauge has automatically turned OFF
OFF and disabled
Flashing amber
Micro-Ion gauge is disabled
ON
Green
• Filament 1 or 2 is ON
• Both filaments are operable
ON
Flashing green
Filament 1 or 2 is inoperable
OFF
Red
Filaments 1 and 2 are inoperable
Table A-3
Micro-Ion Gauge, Gauge LED, and Filament States in Manual Mode
Micro-Ion Gauge State
Gauge LED State
Filament State
OFF automatically
Amber
Micro-Ion gauge has automatically turned OFF
OFF and disabled
Flashing amber
Micro-Ion gauge is disabled
ON
Green
Filament 1 is ON
ON
Flashing green
Filament 2 is ON
OFF
Flashing red
• Filament 1 is inoperable
• Press gauge button to switch to filament 2
OFF
Red
Filaments 1 and 2 are inoperable
Содержание 356 Micro-Ion Plus Series
Страница 2: ......
Страница 24: ...24 Micro Ion Module Instruction Manual 356007 GP...
Страница 58: ...Chapter 3 58 Micro Ion Module Instruction Manual 356007 GP...
Страница 68: ...Appendix A 68 Micro Ion Module Instruction Manual 356007 GP Figure A 1 Automatic Filament Switching Flow Chart...
Страница 70: ...70 Micro Ion Module Instruction Manual 356007 GP Figure A 3 Manual Filament Switching Flow Chart...
Страница 74: ...Index 74 Micro Ion Module Instruction Manual 356007 GP...
Страница 75: ......