DNA Engine Dyad Operations Manual
5-38
• Position the cursor over the
°
C per second field and select.
• Enter 0.5.
• Select <OK>.
You would expect the program step as included in a protocol to appear as fol-
lows:
Incubate at 60.0
°
C for 00:00:30
Ramp to 60
°
C at 0.5
°
C per second
The Increment Temp option
The
Increment Temp option is useful for modifying a temperature step to allow a “per
cycle” increase or decrease of temperature each time the step is executed (see “The
Elements of a Program” near the beginning of this chapter for more information).
Move from the programming window to the Temperature Step window using the <Temp>
button, as before. Enter the appropriate incubation
Temperature (enter 60 as an ex-
ample) and
Time (enter 30 seconds as an example).
• Select
Increment Temp.
• Select <Set Parameters> (for the
Increment Temp option).
The Increment Temperature window appears.
Содержание DNA Engine Dyad PTC-220
Страница 8: ...1 4 DNA Engine Dyad Operations Manual ...
Страница 14: ...2 6 DNA Engine Dyad Operations Manual ...
Страница 38: ...4 18 DNA Engine Dyad Operations Manual ...
Страница 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Страница 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Страница 113: ...9 4 DNA Engine Dyad Operations Manual ...
Страница 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Страница 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Страница 137: ...11 10 DNA Engine Dyad Operations Manual ...
Страница 142: ......
Страница 155: ...Index In 11 service 8 8 set date time 8 3 update soft 8 5 8 6 user name 8 3 V Virtual keyboard 5 18 W Warranties D 1 ...
Страница 156: ...DNA Engine Dyad Operations Manual In 12 ...
Страница 157: ...Declaration of Conformity DoC 1 ...