3-3
Installation
To connect the mouse, please follow these steps:
1. Verify that the Dyad cycler is off. Wait for 10 seconds to ensure that all fans have
stopped rotating.
2. Grasping the sides of the Dyad cycler, tilt the instrument back so that the underside of the
lip is visible.
3. Line up the pins of the mouse connector with the green port and push the connector into
place.
4. Pull the small switch located behind the purple port into the forward position (see figure
2-4). The rear position will activate only the touch pad. The forward position will activate
only the external mouse device. All Dyad cyclers are shipped with the touch pad enabled
(i.e., the switch is in the rear position). Please note that the cycler will recognize either the
touch pad
OR
a mouse, but not both input devices simultaneously.
5. Tip the Dyad cycler back down and power up the system.
Environmental Requirements
Ensure that the area where the DNA Engine Dyad cycler is installed meets the following
conditions, for reasons of safety and performance:
• Nonexplosive environment
• Normal air pressure (altitude below 3000m)
• Ambient temperature 5˚–31˚C
• Relative humidity of 10–90% (noncondensing)
• Unobstructed access to air that is 31˚C or cooler (see below)
• Protection from excessive heat and accidental spills. (Do not place the Dyad cycler near
such heat sources as radiators, and protect it from danger of having water or other fluids
splashed on it, which can cause shorting of its electrical circuits.)
Power Supply Requirements
The DNA Engine Dyad cycler requires 200-240VAC, 50-60Hz, and a grounded outlet on a
minimum 20A line. The Dyad cycler can use voltage in the specified range without adjust-
ment, so there is no voltage-setting switch.
Note:
Do not cut the supplied power cord and attach a different connector. Use a one-piece
molded connector. If required, additional dedicated power cords may be purchased through
MJ Research or authorized distributors.
Содержание DNA Engine Dyad PTC-220
Страница 8: ...1 4 DNA Engine Dyad Operations Manual ...
Страница 14: ...2 6 DNA Engine Dyad Operations Manual ...
Страница 38: ...4 18 DNA Engine Dyad Operations Manual ...
Страница 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Страница 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Страница 113: ...9 4 DNA Engine Dyad Operations Manual ...
Страница 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Страница 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Страница 137: ...11 10 DNA Engine Dyad Operations Manual ...
Страница 142: ......
Страница 155: ...Index In 11 service 8 8 set date time 8 3 update soft 8 5 8 6 user name 8 3 V Virtual keyboard 5 18 W Warranties D 1 ...
Страница 156: ...DNA Engine Dyad Operations Manual In 12 ...
Страница 157: ...Declaration of Conformity DoC 1 ...