Troubleshooting
10-5
Remote
Error Code
Error Code
Reason
Action
7
IFF
The internal fan is not providing
adequate cooling for the
instrument.
Call (888) MJ CYCLE (in the US
or Canada), or your local
distributor (outside the US or
Canada).
22
LSF
An irregularity has been
detected in a heated lid sensor.
The heated lid has been
temporarily turned off. The
sensor will be checked and you
will be notified if any problem is
found.
No action required at this time.
None
LRPI
An imbalance has been
detected in the sample block.
There may be a problem with a
Peltier module.
Call (888) MJ CYCLE (in the US
or Canada), or your local
distributor (outside the US or
Canada).
None
MC
A memory fault has been found.
Stored programs may be
affected.
Call (888) MJ CYCLE (in the US
or Canada), or your local
distributor (outside the US or
Canada).
None
NMA
All available memory has been
filled.
Please delete unused programs,
or call (888) MJ CYCLE (in the
US or Canada), or your local
distributor (outside the US or
Canada).
18
PSF
A possible fault was detected in
the in-sample probe. The
thermal control method has
been switched to Calculated
Mode. Results may have been
affected.
Check probe and connections.
If problem reoccurs please call
(888) MJ CYCLE (in the US or
Canada), or your local distributor
(outside the US or Canada).
9
PSO-CAF
The power supply temperature is
somewhat higher than normal.
The power supply sensor will be
checked and you will be
notified if any problem is found.
Please observe the following
guidelines: 1) place instruments
at least 10cm apart, 2) avoid
placing instruments in any
location where hot air might
enter the intake vents, 3) place
instrument on a hard surface with
no debris or paper underneath,
4) clean all air vents and Alpha
unit fins of dust and debris, 5)
avoid running the instrument in
areas with ambient temperatures
above 25°C. Continue to use the
instrument. If problem reoccurs
please call (888) MJ CYCLE (in
the US or Canada), or your local
distributor (outside the US or
Canada).
Содержание DNA Engine Dyad PTC-220
Страница 8: ...1 4 DNA Engine Dyad Operations Manual ...
Страница 14: ...2 6 DNA Engine Dyad Operations Manual ...
Страница 38: ...4 18 DNA Engine Dyad Operations Manual ...
Страница 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Страница 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Страница 113: ...9 4 DNA Engine Dyad Operations Manual ...
Страница 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Страница 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Страница 137: ...11 10 DNA Engine Dyad Operations Manual ...
Страница 142: ......
Страница 155: ...Index In 11 service 8 8 set date time 8 3 update soft 8 5 8 6 user name 8 3 V Virtual keyboard 5 18 W Warranties D 1 ...
Страница 156: ...DNA Engine Dyad Operations Manual In 12 ...
Страница 157: ...Declaration of Conformity DoC 1 ...