3 Setting Parameters
15
No. 99MBC109B
No.
Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
15
Smoothing
Averages the counter value and then displays it. (This
reduces fluctuation of the lowest-order digit.) You can
specify the number of measurements to average.
●
●
●
○
0: None
1: Display the average of 16 measurements
2: Display the average of 32 measurements
16
Peak value Preset
Sets whether to perform presetting based on the peak
value during Peak mode.
●
●
●
●
0: Disabled
1: Enabled
18
Speed sampling
cycle
(EH-102P/Z/S
only)
In Display mode selection, selects the sampling cycle
when Speed display is selected.
●
●
●
0: 0 ms
2: 100 ms
1: 50 ms
19
Digimatic input
WAIT
Sets the wait time for the Digimatic input signal.
Change this when the Counter cannot read the input
signals from a Digimatic device.
●
0: No WAIT
2: 200 ms WAIT
1: 100 ms WAIT
20
Tolerance output/
BCD output swi-
tchover*
1
Switches between tolerance judgment result output
and BCD output.
●
●
●
●
0: 3-step tolerance
2: BCD output
1: 5-step tolerance
21
BCD output logic*
1
Selects whether to use positive logic (0) or negative
logic (1) for BCD output.
●
●
●
●
0: DATA [L] (sign H)
1: DATA [H] (sign L)
24
RS-232C/USB/
Digimatic output
switchover*
2
Selects which output terminal to use.
●
●
●
●
0: RS-232C
2: Digimatic
1: USB
25
Data transfer
speed*
2
Selects the data transfer speed for RS-232C.
●
●
●
●
0: 4800 bps
2: 19200 bps
1: 9600 bps
26
Parity check*
2
Selects the parity check method for RS-232C.
●
●
●
●
0: None
2: Even numbered
1: Odd numbered
27
Data bit*
2
Selects the length of the data bit for RS-232C.
●
●
●
●
0: 7 bit
1: 8 bit
28
Output trigger*
2
Selects the output trigger method for RS-232C.
●
●
●
●
0: RS-232C command (normal state)
1: RS-232C command (with channel synchronization
function)
2: HOLD trigger
Содержание 542-071
Страница 14: ...xii No 99MBC034B...
Страница 20: ...1 Overview 6 No 99MBC109B MEMO...
Страница 38: ...3 Setting Parameters 24 No 99MBC109B MEMO...