4 Basic Operations
29
No. 99MBC109B
Tips
Each time you press [SEL], the blinking switches between Display A and Display B.
"4.1 Switching Between Display A and B (2-Axis Models Only)" (page 25)
2
Press [MODE].
» Peak mode will be set.
3
Press [A_ZERO]/[ZERO] or [B_ZERO].
» The peak value will be cleared (MAX = MIN = current value, TIR = 0).
Tips
• If the same UNIT (counting method) has been assigned to more than 1 CEL, the peak
value of all of the CELs that are assigned the same UNIT (counting method) will be
cleared when the peak value of one of the CELs is cleared.
• For details about the relationship between BANK, CEL, and UNIT, see "3.4.1 Internal
Structure of the Counter" (page 17).
4.6
Tolerance Value Setting
There are 2 settings for the tolerance value: 3-step and 5-step.
The tolerance values can be set separately for each CEL.
Tips
• Set the value of parameter number 20 to 0 (3-step tolerance) or 1 (5-step tolerance) in
advance.
• For details about I/O output, see "5.5 I/O Connector Terminal Function" (page 42).
4.6.1 3-Step Tolerance Value Setting (3-Step Tolerance
Zone Selection)
With S1 and S4 set as the tolerance values, the 3-step tolerance judgment will
be performed as follows:
Judgment conditions
Tolerance judgment indicator I/O output (PIN number)
Measurement result < S1
Amber indicator on
AL1 (3) / BL1 (11)
S1 ≤ measurement result ≤ S4
Green indicator on
AL3 (5) / BL3 (13)
S4 < measurement result
Red indicator on
AL5 (7) / BL5 (15)
This section explains how to set the 3-step tolerance value.
1
Press and hold [SEL]. (For 1-axis models, proceed to step
2
.)
» The selected Display will blink.
Tips
Each time you press [SEL], the blinking switches between Display A and Display B.
"4.1 Switching Between Display A and B (2-Axis Models Only)" (page 25)