3 Setting Parameters
21
No. 99MBC109B
3.5
CEL-Specific Parameters
Each parameter shown below is set separately for each CEL.
For examples of setting CEL-specific parameters, see "4.7 Optional Constant
Tips
• To set CEL-specific parameters, set 1 as the value for parameter 00. The parameter
number will switch to parameter number 40 for 2-axis models, and to parameter number
41 for 1-axis models.
• The circles in the Counter model columns in the following table indicate whether a
parameter number that is displayed on the Counter is valid or invalid (●: valid, ○: invalid).
No.
Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
40
Individual CEL
display*
1
(2-axis models
only)
Selects the UNIT (counting method) to be displayed
for each CEL when Display mode is set to "Option".
●
●
●
●
0: UNIT A (Count A)
1: UNIT B (Count B)
2: UNIT C (A ± B)
3: UNIT D (Speed A) (except for EH-D)
4: UNIT E (Speed B) (except for EH-D)
5: UNIT F (memory)
41
Calculation with a
constant
Sets whether to multiply the internal counter value
by a predetermined value, by an arbitrary value, or to
not multiply it. The value obtained by multiplying the
internal counter value by the set constant value will be
displayed as the measurement result.
●
●
●
●
0: Do not calculate
2: 2 times
4: Arbitrary value
1: 1/2 times
3: 10 times
42
Hide the lowest-or-
der digit
Hides the lowest-order digit.
However, the lowest-order digit will be included in RS-
232C output and in printouts.
●
●
●
●
0: Display all digits
1: Hide the lowest-order digit
*
1
The Preset value and tolerance value that had been set will be cleared if this setting is changed.