5 External Input/Output Function
46
No. 99MBC109B
5.5.4 Timing Chart
Tolerance judgment result output
-NG
Max. 10 ms
Max. 10 ms
OK
+NG
Upper specification limit
Lower specification limit
Counter data
Tips
• After acquiring the counter data, there is a maximum 10 ms delay before the tolerance
judgment result is output.
• For EH-102D, the length of time until the tolerance judgment result is output after the
counter data enters in the Specification range depends on the connected equipment,
such as the Linear Gage.
BCD output
D1
D2
D8
SET1
BCDCK
READY
Max.
5 ms
Min. 10 ms x 8 = 80 ms
Min.
2 ms
Min.
5 ms
Min.
5 ms
Min.
5 ms
Tips
• "SET1" and "BCDCK" (BCD clock) denote input signals, and "READY" denotes the output
signal.
• The BCD data (D1, D2 in the figure above) will be modified at the fall of the BCD clock
while SET1 is input. When the BCD data is accepted, the READY signal falls. The time to
accept the data after the fall of the BCD clock is 5 ms maximum. The READY signal rises
together with the BCD clock.