3 Setting Parameters
12
No. 99MBC109B
5
Press [P.SET].
» The current value of parameter
number 12 of INPUT B will be
displayed.
Tips
For 1-axis models, the parameter number will advance. Proceed to step
8
.
6
Repeatedly press [A_ZERO]/[ZERO] to set the set value to 1 (resolution:
5 µm).
» The value of INPUT B will be set to 1
(the Linear Gage resolution of INPUT
B will be set to 5 µm).
Parameter number INPUT
number
Set value
7
Press [P.SET].
» The set value will be applied.
8
Press and hold [P.SET], and then press [A_ZERO]/[ZERO].
» The Counter will return to the stand-
by state.
3.2
Basic Parameters
This section explains the basic parameters related to measuring. Be sure to set
these settings before measuring.
Tips
• Correct measurement results may not be obtained if the settings are incorrect.
• The circles in the Counter model columns in the following table indicate whether a
parameter number that is displayed on the Counter is valid or invalid (●: valid, ○: invalid).
No.
Setting item
Description/Allowable values
(the values in bold indicate the default value)
Counter model
P Z S D
00
Parameter mode
Used for setting/saving/loading parameters or setting
the optional constant value.
●
●
●
●
0: Parameter setting
1: CEL-specific parameter setting
2: Constant value setting
3: Save parameters
4: Load parameters
Содержание 542-071