5 External Input/Output Function
48
No. 99MBC109B
HOLD/Error reset
Min.
10 ms
Min. 10 ms
Max. 15 ms
HOLD
BCD data
Error cancellation
(For EH-102Z: origin detection wait state)
Data latch
10 µs (10 ms)
Tips
• The data latch function and error cancellation function are assigned to the HOLD signal.
• The data latch function will be executed 10 µs (or 10 ms when the HOLD input is active
axis by axis) after the fall of the HOLD signal. The error cancellation will be executed with
the rising HOLD signal.
• When outputting BCD data, start the data output within 15 ms after the data latch.
• For EH-102D, the length of time until the data latch depends on the connected equipment,
such as a Linear Gage.
• During HOLD input, the UNIT indicator will blink.