135
Data Processor
Applications
<Setting the Multi-calibration>
This sets the multi-calibration function which corrects indicated values using both white calibration
and user calibration data.
When setting or canceling multi-calibration, you must set the color difference target color
again.
The color difference target color is not corrected using user calibration performed after mea-
surement.
•
See P.145 for details on multi-calibration.
•
Multi-calibration cannot be set if white calibration and user calibration are not done ahead of time.
[Operating Procedure]
1
Press the Option key
Option
4
GHI
while in the measurement
screen.
•
The option selection screen appears.
2
Select “MALTI CAL.” using the
key.
Selection screen
[OPTION]
TOLERANCE
AUTO MEAS.
DATE & TIME
IMPORT
MALTI CAL.
P00
C P
[0001]
[OPTION]
TOLERANCE
AUTO MEAS.
DATE & TIME
IMPORT
MALTI CAL.
P00
C P
[0001]
2,4
3,5
1