134
Data Processor
Applications
User Calibration
[CALIB.]
◎USER CALIB.
Y= 80.72
x=0.3208
y=0.2121
◎CAL. Ch NAME A
MINOLTA
P00
C P
[0001]
[CALIB.]
◎USER CALIB.
Y= 80.72
x=0.3208
y=0.2121
◎CAL. Ch NAME A
Ch03
P00
C P
[0001]
8
Move the cursor using the
key and set the value
using the numeric pad.
Input range: 0.01
≤
X, Y, Z
≤
160
9
Place the measuring head vertically above a target
specimen.
10
Press the Measure/Enter key
(or measurement
button on the measuring head) after making sure the
ready lamp is ON.
Do not move the measuring head during measurement.
11
Once measurement is complete, set the calibration channel name using the
key and the numeric pad.
(How to set the calibration channel name)
1) Select input mode using the
key.
•
Pressing the
key changes the input modes in the following order: upper case Latin
letters
→
numbers
→
lower case Latin letters.
2)
Move the cursor using the
key and input the color difference target color name using
the numeric pad.
•
The name is up to 9 characters (see P.20 “As input keys” for details.)
12
Once settings are complete, press the Measure/Enter key
.
•
Settings are complete, complete sound beeps and the display returns to the measurement screen.
Setting screen
Setting screen
POWER
Содержание CR-400 -
Страница 23: ...21 Chapter 1 Measuring Head 1 This chapter describes how to use the measuring head alone...
Страница 54: ...52...
Страница 59: ...57 Chapter 3 Data Processor This chapter describes how to use connected the Measuring Head and Data Processor...
Страница 143: ...141 Chapter 4 Description...
Страница 156: ...154 Description Dimensions Measuring Head Unit mm CR 400 Head 217 63 102 40 2 32...
Страница 157: ...155 Description Unit mm CR 410 Head 67 5 244 63 102 32...
Страница 158: ...156 Description Data Processor Unit mm DP 400 Dimensions 100 255 73...