139
Data Processor
Applications
<Deleting selected calibration channel>
The user calibration channel can be deleted as follows.
[Operating Procedure]
1
Press the Calibration key
while in the measure-
ment screen.
•
The calibration screen appears.
2
Select “USER CALIB.” using the
key.
3
Press the Measure/Enter key
.
•
The user calibration list screen appears.
Calibration screen
List screen
Calibration screen
[CALIB.]
◎WHITE CALIB.
Y= 9 3 .5 0
x=0.3 1 1 4
y=0.3 1 9 0
USER CALIB.
P00
C P
[0001]
[CALIB.]
◎WHITE CALIB.
Y= 9 3 .5
x=0.3 1 1 4
y=0.3 1 9 0
USER CALIB.
P00
C P
[0000]
[CALIB.] 1/1
◎USER CALIB.
NEW
Ch02
Ch01
P00
C P
[0001]
2,4,6
3,7,8
1
5