86
Data Processor
Measurement
When using the instrument for long periods of time, the displayed value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend
that white calibration is done regularly using the white calibration plate. White calibration should also
be done before measuring after a long time has passed since the instrument was last used.
White calibration should be done under the same temperature conditions as measurement.
[Operating Procedure]
1
Press the Calibration key
while in the measurement screen.
•
The white calibration screen appears.
2
Set the calibration data listed on the back of the white calibration plate cover
using the
key and the numeric pad.
•
The cursor moves automatically to the next position when setting data using the numeric pad.
•
Use the
to move the cursor to make corrections.
•
Go to step 3, if white calibration data has already been set.
[Example]
•
Illuminants C
Y=93.5 x=0.3114 y=0.3190
•
Illuminants D
65
Y=93.5 x=0.3140 y=0.3318
POWER
1
Ready lamp
White calibration plate
2
4
Color Space key
4
3
White Calibration
No white calibration
data has been set.
White calibration data
has been set.
[CALIB.]
◎WHITE CALIB.
Y= 0.00
x=0.0000
y=0.0000
USER CALIB.
P00
C P
[0000]
[CALIB.]
◎WHITE CALIB.
Y= 9 3 .5
x=0.3 1 1 4
y=0.3 1 9 0
USER CALIB.
P00
C P
[0000]
Содержание CR-400 -
Страница 23: ...21 Chapter 1 Measuring Head 1 This chapter describes how to use the measuring head alone...
Страница 54: ...52...
Страница 59: ...57 Chapter 3 Data Processor This chapter describes how to use connected the Measuring Head and Data Processor...
Страница 143: ...141 Chapter 4 Description...
Страница 156: ...154 Description Dimensions Measuring Head Unit mm CR 400 Head 217 63 102 40 2 32...
Страница 157: ...155 Description Unit mm CR 410 Head 67 5 244 63 102 32...
Страница 158: ...156 Description Data Processor Unit mm DP 400 Dimensions 100 255 73...