42
9-1-2.
Sensor Channels
*To edit values or names, double-click the corresponding item name.
Press the [ENTER] or [TAB] key after editing to enter the changes made.
1) Operation buttons
Operation
button name
Description
Copy CH1 to all
CH
Pressing this button refreshes all of the channels to reflect the setting
details for CH1.
Note that the CH names remain unchanged.
Setting Description
Enable
Check only those channels to be used.
Waveform data will be acquired and plotted only for the channels
enabled.
Rated capacity
Sets the rated capacity for the sensor used.
Can be selected from 196.1 N (20K), 490.3 N (50K),
980.7 N (100K), 1.961 kN (200K), 4.903 kN (500K),
9.807 kN (1T), 29.42 kN (3T). Double-click to enter a
numerical value directly as desired. The units will be
"N" for displayed values and numerical input values,
and "kN" cannot be entered.
* Selections 200K, 500K, 1T, 3T, and numerical input
values exceeding 2000.0 N are possible with ROM
Ver.1.200 or later.
Rated output
Sets the rated output for the sensor used.*
Shape
Sets the pressure transmitting pin shape to [Round], [Rectangle], or
[Other].
Diameter
Enter the dimensions if the pressure transmitting pin shape is set to
[Round].
Width
Enter the dimensions if the pressure transmitting pin shape is set to
[Rectangle].
Height
Enter the dimensions if the pressure transmitting pin shape is set to
[Rectangle].
Area
Enter the cross-sectional area if the pressure transmitting pin shape is
set to [Other].
Name
Sets the channel name used on the [Measurement data display] and
[Display channel selection screen] for the Measurement screen;
[Information screen] and [Statistics screen] for the Data processing
screen; and data output displays.
Содержание MPC-201-25
Страница 1: ...Molding Pressure Conversion Module MPC 201 25 Instruction Manual EN294 1697G ...
Страница 2: ......
Страница 14: ...XII ...
Страница 18: ...XVI 24 4 Vertically positioning the ejector plate 102 24 5 Removing a sensor 103 ...
Страница 33: ...15 3 Right click to display the context menu and select Properties ...
Страница 34: ...16 4 Click on Update Driver in the Driver tab 5 Select Browse my computer for driver software ...
Страница 43: ...25 5 Overlay screen 6 Trend Peak Pressure screen 7 Trend Integral screen ...
Страница 54: ...36 Reset scale 1 Right click on the graph to be reset to its original scale 2 Click to select Reset scale ...
Страница 122: ......
Страница 123: ......