22. Hardware Specifications
22-1.
Analog Specifications
Input spe
cifications
Bridge power supply
GV 5 V ±0.25 V, max. 20 mA (per channel)
Applicable transducers
LSMS-20K-S06, LSMS-50K-S06, LSMS-100K-S06
LSMS-200K-S06, LSMS-500K-S06, LSMS-1T-S06
LSMS-3T-S06
* 200K, 500K, 1T, and 3T (2000.0 N or more)
are available with ROM Ver.1.200 or later.
No. of inputs
8 channels
Input range
-0.1 mV/V
~
0.8 mV/V
Zero adjustment range
Adjustable between -0.1 mV/V
~
0.7 mV/V using zero set
function
Accuracy ±2.0%F.S.
Sampling frequency
10 ms per channel
(can be set to 0.5, 1, 2, 5, 20, 50 ms)
Sampling interval
Max. 120 s (Max. 600 s for long time measurement)
A/D internal resolution
16 bit
Analog filter
Approx. 500 Hz
Output sp
ecifi
cation
s
No. of outputs
8 channels
Output voltage range
0 V
~
10 V DC
Load resistance
Min. 5 k
Ω
Resolution Min.
1/12000
Output cycles
Synchronized with A/D sampling
Over-range
Approx. 11 V when exc110 % of rated pressure
Approx. -1 V when below -10 % of rated pressure
22-2.
Digital Specifications
Measurement range
0 MPa
~
999.99 MPa (based on sensor area)
Units MPa
Status LEDs
POWER, ALARM, ERROR
Saving
EEPROM
Measurement condition settings, alarm function settings
RAM
Zero set data, measurement mode (overlay or continuous)
22-3.
Interface
External control
input
Start trigger
Zero-voltage contact input
Enabled once with pulse width of at least 50 ms
Alarm cancel
External control
output
Alarm
Open collector output
VCE = DC35 Vmax., Ic = DC100 mAmax
USB
Specifications
USB 2.0/1.1 compatible
Output connector
miniUSB connector B type (female)
Содержание MPC-201-25
Страница 1: ...Molding Pressure Conversion Module MPC 201 25 Instruction Manual EN294 1697G ...
Страница 2: ......
Страница 14: ...XII ...
Страница 18: ...XVI 24 4 Vertically positioning the ejector plate 102 24 5 Removing a sensor 103 ...
Страница 33: ...15 3 Right click to display the context menu and select Properties ...
Страница 34: ...16 4 Click on Update Driver in the Driver tab 5 Select Browse my computer for driver software ...
Страница 43: ...25 5 Overlay screen 6 Trend Peak Pressure screen 7 Trend Integral screen ...
Страница 54: ...36 Reset scale 1 Right click on the graph to be reset to its original scale 2 Click to select Reset scale ...
Страница 122: ......
Страница 123: ......