ZS3 Instructions for Use
123
|
P a g e
•
Do not use other disinfection methods like Iodine, Steam, Heat or Ethylene Oxide.
•
When servicing the ZS3 System, always be sure to turn the circuit breaker to the OFF position.
Precautions
•
Transducers are individually licensed according to system configuration. Only the following
transducers are available without an advanced feature configuration: L30-8, P7-3c, C12-4lp, C4-1,
C6-2, C6-1, C9-3, C9- 3sp, C8-3 3D, C10-3, C18-5, E9-3 3D, E9-3, E9-4, L8-3, L10-5, L14-5w,
L14-5sp, L20-5, P4-1c,
•
P8-3TEE and P8-3mTEE. The A2CW and A5CW transducers are enabled if the system is
configured with either the Echocardiography or the Advanced Vascular Imaging Feature Sets.
Attempting to use any other transducers will result in an error message.
•
Be aware of the potential hazards associated with the environment where the ultrasound systems
will be used. The systems and/or the external equipment can be damaged if signal levels are not
appropriate. If peripheral equipment not specifically authorized by Mindray is to be connected to
the system, it must meet all applicable electrical safety standards that apply to the system in order
to maintain safety integrity. Any equipment not supplied by Mindray must be approved by Mindray.
•
Use of non-Mindray-approved equipment may result in an unsafe condition, impair operation of
the ultrasound system, impair diagnostic capabilities, and void your warranty or service contract
coverage.
•
Compatible transducers have a specific range of acceptable application use. Users are advised to
restrict each transducer’s use to those applications. Excessive bending, twisting, pulling, dragging,
or compression of transducer cables may cause failure or intermittent operation of the system.
•
Avoid rolling the cart wheels over cables. Use of cable hooks is recommended to minimize chance
of damage to cables.
•
If a transducer that has not been approved and licensed for use with the ultrasound system to
which it is connected, or if a licensed transducer is not properly connected, the corresponding
transducer identification data will not display on the LCD Display. Imaging will be disabled.
Resolve this issue before continuing use.
•
The ultrasound system and/or the transducer could be damaged if an unapproved transducer is
connected.
•
Improper cleaning or disinfection of patient applied parts may cause permanent damage.
•
Carefully review the manufacturer’s directions for any component used with the ultrasound
system. Mindray assumes no liability with respect to single-use devices that are reused,
reprocessed, or resterilized and makes no warranties, expressed or implied (including the
warranties of merchantability or fitness for a particular use), with respect to such devices.
•
Use of peripherals or other equipment not provided by Mindray may result in system damage or
degraded performance. Carefully review the labeling of any such equipment before connecting to
the ultrasound system.
•
Improper setting of imaging controls may obscure diagnostically valuable information in the
display. The factory default preset maps were preselected as appropriate for most imaging
circumstances. Improper user configuration of custom presets may obscure diagnostically
valuable information.
•
Mindray ultrasound system is manufactured in compliance with existing electromagnetic immunity
(EMI) and electromagnetic compatibility (EMC) requirements. Use of the systems in the presence
of an electromagnetic field can cause degradation of the ultrasound image.
•
Electrostatic discharge (ESD), or static shock, is a naturally occurring phenomenon. ESD is
common in low humidity, which can be caused by heating or air conditioning. ESD shock occurs
when electrical energy is discharged from one body, to a differently charged body. To lessen the
occurrence of ESD, use antistatic spray on carpets and flooring, and antistatic mats.
•
To minimize potential ESD damage and electrical contact contamination, avoid touching the metal
contacts for the transducer connections, at both the ZS3 port and the transducer connector.
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...