ZS3 Instructions for Use
140
|
P a g e
TABLE 3
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT-
GUIDANCE
Conduced RF IEC
61000-4-6
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur
radio bands between
0,15 MHz and 80 MHz
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur
radio bands
between
0,15 MHz and 80
MHz
Portable and mobile RF
communications equipment should
be used no closer to any part of
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter. Recommended
separation distance
d = 1.2 x
P
d = 1.2 x
P
80 MHz to 800 MHz
d = 2.3 x
P
800 MHz to 2.5GHz
Where, P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Radiated RF IEC
61000-4-3
10 V/m
80MHz - 2.7GHz
10 V/m
80MHz - 2.7GHz
Proximity fields
from RF
wireless
communications
equipment
IEC 61000-4-3
27 V/m
380–390 MHz
27 V/m
28 V/m
430–470 MHz, 800–960
MHz, 1700–1990 MHz,
2400–2570 MHz
28 V/m
9 V/m
704–787 MHz, 5100–
5800 MHz
9 V/m
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Содержание Zonare ZS3
Страница 1: ...I ZS3 Ultrasound System Instructions for Use ...
Страница 6: ......
Страница 12: ......
Страница 38: ......
Страница 56: ...ZS3 Instructions for Use 48 P a g e Figure 4 6 P8 3m Adult and Pediatric TEE Scan Plane adjustment ...
Страница 74: ......
Страница 78: ......
Страница 86: ......
Страница 110: ......
Страница 128: ......
Страница 146: ......
Страница 150: ......
Страница 158: ......
Страница 160: ......
Страница 164: ...ZS3 Instructions for Use 147 P a g e ...
Страница 165: ...P N 046 012347 01 3 0 ...